Purpose: Novel, preclinical radiotherapy modalities are being developed at synchrotrons around the world, most notably stereotactic synchrotron radiation therapy and microbeam radiotherapy at the European Synchrotron Radiation Facility in Grenoble, France. The imaging and medical beamline (IMBL) at the Australian Synchrotron has recently become available for preclinical radiotherapy and imaging research with clinical trials, a distinct possibility in the coming years. The aim of this present study was to accurately characterize the synchrotron-generated x-ray beam for the purposes of air kerma-based absolute dosimetry.
Methods: The authors used a theoretical model of the energy spectrum from the wiggler source and validated this model by comparing the transmission through copper absorbers (0.1-3.0 mm) against real measurements conducted at the beamline. The authors used a low energy free air ionization chamber (LEFAC) from the Australian Radiation Protection and Nuclear Safety Agency and a commercially available free air chamber (ADC-105) for the measurements. The dimensions of these two chambers are different from one another requiring careful consideration of correction factors.
Results: Measured and calculated half value layer (HVL) and air kerma rates differed by less than 3% for the LEFAC when the ion chamber readings were corrected for electron energy loss and ion recombination. The agreement between measured and predicted air kerma rates was less satisfactory for the ADC-105 chamber, however. The LEFAC and ADC measurements produced a first half value layer of 0.405 ± 0.015 and 0.412 ± 0.016 mm Cu, respectively, compared to the theoretical prediction of 0.427 ± 0.012 mm Cu. The theoretical model based upon a spectrum calculator derived a mean beam energy of 61.4 keV with a first half value layer of approximately 30 mm in water.
Conclusions: The authors showed in this study their ability to verify the predicted air kerma rate and x-ray attenuation curve on the IMBL using a simple experimental method, namely, HVL measurements. The HVL measurements strongly supports the x-ray beam spectrum, which in turn has a profound effect on x-ray dosimetry.
Download full-text PDF |
Source |
---|---|
http://dx.doi.org/10.1118/1.4803675 | DOI Listing |
Med Phys
January 2025
Department of Medical Physics, School of Medicine and Public Health, University of Wisconsin-Madison, Madison, Wisconsin, USA.
Background: A cylindrical free-air chamber, the Attix FAC, is used for absolute air-kerma measurements of low-energy photon beams at the University of Wisconsin Medical Radiation Research Center. Correction factors for air-kerma measurements of specific beams were determined in the 1990s. In order to measure air-kerma rates of beams in development, new correction factors must be computed.
View Article and Find Full Text PDFIndian Heart J
January 2025
Department of Cardiology, Sri Jayadeva Institute of Cardiovascular Sciences and Research, Bangalore, India.
Background: Cardiac catheterization via the trans-radial approach (TRA) has shown several advantages over the trans-femoral approach (TFA) but with a concern of higher radiation exposure. Considering the growing experience with TRA, this study compares patient's radiation during coronary angiography using TRA versus TFA.
Methods: This study included consecutive patients undergoing coronary angiogram over a year at tertiary hospital performed by experienced operators through radial or femoral access.
Heliyon
January 2025
Faculty of Medicine and Health Sciences, Sana'a University, Sana'a, Yemen.
Background: Dynamic Coronary Roadmap (DCR) is a new PCI method that may reduce contrast dose and contrast-associated acute kidney injury (CA-AKI) risk. This paper evaluates DCR-guided PCI versus standard angiography PCI for contrast usage, procedure time, and CA-AKI risk.
Methods: On May 1, 2024, we searched PubMed, Scopus, Embase, Cochrane Library, and clinicaltrials.
J Vasc Surg Cases Innov Tech
April 2025
Department of Radiology and Nuclear Medicine, University Teaching and Research Hospital Lucerne, Luzern, Switzerland.
Objective: The aim of this single-center case series is to demonstrate that an ultra-low dose (ULD) can be routinely achieved in the hybrid operating room in standard endovascular aortic repair (EVAR) for infrarenal abdominal aortic aneurysm by adjusting the manufacturer's predefined imaging parameters, hardware configurations and user protocols (including benchmarking).
Methods: The hybrid operating room manufacturer predefined EVAR software setup of the dose exposure control software (OPTIQ, Siemens Healthineers, Forchheim, Germany) at our university medical center was screened for possible improvements regarding radiation dose application. Tests on a water-equivalent as well as polymethyl methacrylate phantom model to assess the impact of technical settings were performed, including comparison of settings for exposure control software, different magnification, collimation configurations and detector distance.
Med Phys
January 2025
Department of Medical Imaging, Radboud University Medical Center, Nijmegen, The Netherlands.
Background: Dedicated breast computed tomography (bCT) systems offer detailed imaging for breast cancer diagnosis and treatment. As new bCT generations are developed, it is important to evaluate their imaging performance and dose efficiency to understand differences over previous models.
Purpose: To characterize the imaging performance and dose efficiency of a second-generation (GEN2) bCT system and compare them to those of a first-generation (GEN1) system.
Enter search terms and have AI summaries delivered each week - change queries or unsubscribe any time!