Purpose: To investigate the effect of energy (kVp) and filters (no filter, half Bowtie, and full Bowtie) on the dose response curves of the Gafchromic XRQA2 film and nanoDot optical stimulated luminescence dosimeters (OSLDs) in CBCT dose fields. To measure surface and internal doses received during x-ray volume imager (XVI) (Version R4.5) and on board imager (OBI) (Version 1.5) CBCT imaging protocols using these two types of dosimeters.

Methods: Gafchromic XRQA2 film and nanoDot OSLD dose response curves were generated at different kV imaging settings used by XVI (software version R4.5) and OBI (software version 1.5) CBCT systems. The settings for the XVI system were: 100 kVp∕F0 (no filter), 120 kVp∕F0, and 120 kVp∕F1 (Bowtie filter), and for the OBI system were: 100 kVp∕full fan, 125 kVp∕full fan, and 125 kVp∕half fan. XRQA2 film was calibrated in air to air kerma levels between 0 and 11 cGy and scanned using reflection scanning mode with the Epson Expression 10000 XL flat-bed document scanner. NanoDot OSLDs were calibrated on phantom to surface dose levels between 0 and 14 cGy and read using the inLight(TM) MicroStar reader. Both dosimeters were used to measure in field surface and internal doses in a male Alderson Rando Phantom.

Results: Dose response curves of XRQA2 film and nanoDot OSLDs at different XVI and OBI CBCT settings were reported. For XVI system, the surface dose ranged between 0.02 cGy in head region during fast head and neck scan and 4.99 cGy in the chest region during symmetry scan. On the other hand, the internal dose ranged between 0.02 cGy in the head region during fast head and neck scan and 3.17 cGy in the chest region during chest M20 scan. The average (internal and external) dose ranged between 0.05 cGy in the head region during fast head and neck scan and 2.41 cGy in the chest region during chest M20 scan. For OBI system, the surface dose ranged between 0.19 cGy in head region during head scan and 4.55 cGy in the pelvis region during spot light scan. However, the internal dose ranged between 0.47 cGy in the head region during head scan and 5.55 cGy in the pelvis region during spot light scan. The average (internal and external) dose ranged between 0.45 cGy in the head region during head scan and 3.59 cGy in the pelvis region during spot light scan. Both Gafchromic XRQA2 film and nanoDot OSLDs gave close estimation of dose (within uncertainties) in many cases. Though, discrepancies of up to 20%-30% were observed in some cases.

Conclusions: Dose response curves of Gafchromic XRQA2 film and nanoDot OSLDs indicated that the dose responses of these two dosimeters were different even at the same photon energy when different filters were used. Uncertainty levels of both dosimetry systems were below 6% at doses above 1 cGy. Both dosimetry systems gave almost similar estimation of doses (within uncertainties) in many cases, with exceptions of some cases when the discrepancy was around 20%-30%. New versions of the CBCT systems (investigated in this study) resulted in lower imaging doses compared with doses reported on earlier versions in previous studies.

Download full-text PDF

Source
http://dx.doi.org/10.1118/1.4803466DOI Listing

Publication Analysis

Top Keywords

xrqa2 film
28
film nanodot
24
dose ranged
24
cgy head
24
head region
24
gafchromic xrqa2
20
dose response
16
response curves
16
nanodot oslds
16
cgy
15

Similar Publications

In recent years, radiochromic films have begun to be used for dosimetry in mammography; however, the most sensitive GAFCHROMIC XR-QA2 (XR-QA2) film is no longer available owing to its discontinuation. In this study, we evaluated the sensitivity characteristics of GAFCHROMIC LD-V1 (LD-V1) as an alternative to XR-QA2 in the field of mammography, at a low dose and low energy. Our results show that the average ratio of the concentration change of LD-V1 divided by the concentration change of XR-QA2 at each absorbed dose was 53.

View Article and Find Full Text PDF

Purpose: This study aims to compare two methods for the organ dose evaluation in computed tomography (CT) in the head- and thorax regions: an experimental method, using radiochromic films, and a computational one, using a commercial software.

Methods: Gafchromic® XR-QA2 and EBT-3 were characterized in terms of energetic, angular, and irradiation configurations dependence. Two free-in-air irradiation calibration configurations were employed using a CT scanner: with the sensitive surface of the film orthogonal (OC) and parallel (PC) to the beam axis.

View Article and Find Full Text PDF

During a single scan using computed tomography, an X-ray tube orbits along a 360°-circular path around the patient. A scan obtained using the half-cylindrical type phantoms with a radiochromic film sandwiched in between reveals a pixel value map illustrating the two-dimensional (2D) dose distribution. A three-dimensional (3D) dose distribution can be obtained with a 360° rotation of the 2D dose map.

View Article and Find Full Text PDF

The purpose of this study was to evaluate the performance of solid-state sensor based computed tomography dose profiler (CTDP) probe for measurement of standard computed tomography dose metric CTDIand free in air geometric efficiency for various beam widths available in a 128-slice CT scanner and also to estimate the efficiency of CTDImetric. The response accuracy of CTDP probe was verified using a standard 100 mm long ionization chamber. The geometric efficiency measurements performed by the CTDP probe were validated using XR-QA2 radiochromic film measurements.

View Article and Find Full Text PDF

Purpose: In this study, we used Gafchromic™ film XR-QA2 and RT-QA2 to characterize the film energy response against various radionuclides. We introduce a neutron depletion theoretical model that can describe film response as a function of cumulated activity. The film response was investigated with respect to different backscatter media such as polystyrene, perspex, lead and corrugated fibreboard carton (CFC).

View Article and Find Full Text PDF

Want AI Summaries of new PubMed Abstracts delivered to your In-box?

Enter search terms and have AI summaries delivered each week - change queries or unsubscribe any time!