FK506 (tacrolimus), a macrolide immunosuppressant, is widely used in pediatric transplant patients, but a relatively narrow therapeutic window in children vs adults requires close and accurate monitoring of whole blood FK506 levels. High-pressure liquid chromatography/tandem mass spectrometry (HPLC/MS/MS)-based assays have been viewed as the gold standard but are more time and labor intensive than cloned enzyme donor immunoassay (CEDIA). To analyze differences between the 2 assays, we assayed FK506 in 348 split samples simultaneously by both methods. A further 70 samples were stratified by organ transplantation type: cardiac (13%), renal (23%), small bowel (22%), or liver transplantation (42%). Results were analyzed using standard statistical techniques for method comparison. CEDIA overestimated the FK506 value relative to HPLC/MS/MS by more than 20% in 40% of cases (139/348), whereas CEDIA underestimated the FK506 value relative to HPLC/MS/MS by more than 20% in 13.5% of cases, for a total inaccuracy of 53% using a ±20% cutoff. Only 28% of samples (99/348) measured by CEDIA were within 10% of the value obtained by HPLC/MS/MS. Bland-Altman analysis showed a mean bias of 9.5% in favor of CEDIA over HPLC/MS/MS (95% confidence interval, 6.1%-12.9%). Positive bias was greatest for liver transplant and R(2) values were lowest for intestinal transplant patients, indicating that HPLC/MS/MS may be a better option for this pediatric transplant subgroup.

Download full-text PDF

Source
http://dx.doi.org/10.1309/AJCPB6EUC5QHLGNVDOI Listing

Publication Analysis

Top Keywords

comparison cedia
8
pediatric transplant
8
transplant patients
8
fk506 relative
8
relative hplc/ms/ms
8
hplc/ms/ms 20%
8
fk506
6
hplc/ms/ms
6
cedia
5
cedia fk506
4

Similar Publications

Degradation of PET Bottles by an Engineered PETase.

Polymers (Basel)

April 2023

Facultad de Ciencia e Ingeniería en Alimentos y Biotecnología, Universidad Técnica de Ambato, Ambato 180216, Ecuador.

Extensive plastic production has become a serious environmental and health problem due to the lack of efficient treatment of plastic waste. Polyethylene terephthalate (PET) is one of the most used polymers and is accumulating in landfills or elsewhere in nature at alarming rates. In recent years, enzymatic degradation of PET by PETase (PETase), a cutinase-like enzyme, has emerged as a promising strategy to completely depolymerize this polymer into its building blocks.

View Article and Find Full Text PDF

Biosurfactant from DS03: Properties and Application in Cleaning Out Place System in a Pilot Sausages Processing.

Microorganisms

July 2022

Escuela Superior Politécnica del Litoral, ESPOL, Facultad de Ingeniería Mecánica y Ciencias de la Producción, Campus Gustavo Galindo, Km 30.5, Vía Perimetral, Guayaquil P.O. Box 09-01-5863, Ecuador.

Biosurfactants (BS) are amphiphilic molecules that align at the interface reducing the surface tension. BS production is developed as an alternative to synthetic surfactants because they are biodegradable, with low toxicity and high specificity. BS are versatile, and this research proposes using a biosurfactant crude extract (BCE) as part of cleaning products.

View Article and Find Full Text PDF

Urine toxicology screening by liquid chromatography time-of-flight mass spectrometry in a quaternary hospital setting.

Clin Biochem

September 2021

Clinical Biochemistry, Alfred Health, 55 Commercial Rd, Melbourne 3004, Australia; School of Public Health and Preventative Medicine, Commercial Rd, Melbourne 3004, Monash University, Australia.

Objective: Validation of a non-targeted method for urine drug screening (UDS) by liquid chromatography quadrupole time-of-flight mass spectrometry (LC-QTOF), and comparison to an established GC-MS method in a hospital setting.

Methods: 217 UDS specimens sent to a quaternary hospital pathology department, were analysed by a CEDIA® immunoassay screen (six drug panels; amphetamines, barbiturates, benzodiazepines, cocaine metabolites, cannabinoids and opiates) on an Abbott Architect instrument. Specimens were subsequently analysed by an established non-targeted qualitative GC-MS method and results compared with a general unknown screening method by LC-QTOF that was under evaluation as a replacement method.

View Article and Find Full Text PDF

Lam. leaf powder antioxidant activity and cytotoxicity in human primary fibroblasts.

Nat Prod Res

December 2021

Chemical Engineering Department, Engineering, Applied Sciences & Simulation Group (GICAS), Colegio de Ciencias e Ingenierías, Universidad San Francisco de Quito USFQ, Quito, Ecuador.

Lam. (MO) leaf powder has been well studied, however, understanding how extraction methods of antioxidant compounds affect human primary fibroblasts still needs to be determined. The antioxidant capacity was analyzed through a copper reduction capacity method and primary human skin fibroblasts were evaluated for cytotoxicity using MTT (3-(4,5-dimethylthiazol-2-yl)-2,5-diphenyltetrazolium bromide) assay.

View Article and Find Full Text PDF

Comparison of several immunoassays used in drugs of abuse screening: Assessment against gold standard methods and calculation of measurement uncertainty.

J Pharmacol Toxicol Methods

November 2020

NSW Health Pathology, Forensic & Analytical Science Service (FASS), Sydney, Australia; Faculty of Medicine and Health, Sydney University, Australia. Electronic address:

Background: Immunoassays provide simple, powerful and inexpensive screening methods for urine drug screening. Other substances and/or factors may interfere with the test and cause false or positive results. It is essential to understand the differences between methods to be able to evaluate their impact on the results.