AI Article Synopsis

  • - The study investigates how different implant angulations (0°, 17°, and 30°) and connection types (external hexagon and tapered) affect tension distribution on bone tissue around dental implants using 3D finite element methods.
  • - Twelve configurations were simulated, measuring stress in the bone at various regions under axial and oblique loads, revealing that the external hexagon connection resulted in greater stress concentrations compared to the tapered connection.
  • - Results indicated that oblique loading was more harmful to bone tissue, particularly with external hexagon implants, and highlighted that the buccal region experienced the highest stress concentrations.

Article Abstract

Background: The purpose of this study is to analyze the tension distribution on bone tissue around implants with different angulations (0°, 17°, and 30°) and connections (external hexagon and tapered) through the use of three-dimensional finite element and statistical analyses.

Methods: Twelve different configurations of three-dimensional finite element models, including three inclinations of the implants (0°, 17°, and 30°), two connections (an external hexagon and a tapered), and two load applications (axial and oblique), were simulated. The maximum principal stress values for cortical bone were measured at the mesial, distal, buccal, and lingual regions around the implant for each analyzed situation, totaling 48 groups. Loads of 200 and 100 N were applied at the occlusal surface in the axial and oblique directions, respectively. Maximum principal stress values were measured at the bone crest and statistically analyzed using analysis of variance. Stress patterns in the bone tissue around the implant were analyzed qualitatively.

Results: The results demonstrated that under the oblique loading process, the external hexagon connection showed significantly higher stress concentrations in the bone tissue (P <0.05) compared with the tapered connection. Moreover, the buccal and mesial regions of the cortical bone concentrated significantly higher stress (P <0.005) to the external hexagon implant type. Under the oblique loading direction, the increased external hexagon implant angulation induced a significantly higher stress concentration (P = 0.045).

Conclusions: The study results show that: 1) the oblique load was more damaging to bone tissue, mainly when associated with external hexagon implants; and 2) there was a higher stress concentration on the buccal region in comparison to all other regions under oblique load.

Download full-text PDF

Source
http://dx.doi.org/10.1902/jop.2013.120713DOI Listing

Publication Analysis

Top Keywords

external hexagon
16
three-dimensional finite
12
finite element
12
bone tissue
12
0° 17°
8
17° 30°
8
30° connections
8
connections external
8
hexagon tapered
8
axial oblique
8