Background Context: There is no universal consensus regarding the biomechanical aspects and relevance on the primary stability of misplaced pedicle screws.

Purpose: The study is aimed to the determination of the correlation between axial pullout forces of pedicle screws with the possible screw misplacement, including mild and severe cortical violations.

Methods: Eighty-eight monoaxial pedicle screws were implanted into 44 porcine lumbar vertebral bodies, paying attention on trying to obtain a wide range of placement accuracy. After screw implantation, all specimens underwent a spiral computed tomography scan, and the screw placements were graded following the scales of Laine et al. and Abul Kasim et al. Axial pullout tests were then performed on a servohydraulic material testing system.

Results: Decreasing pullout forces were determined for screws implanted with increasing cortical violation. A smaller influence of cortical violations in the medial direction with respect to the lateral direction was observed. Screws implanted with a large cortical violation and misplacement in the craniocaudal direction were found to be significantly less stable than screws having comparable cortical violation but in a centered sagittal position.

Conclusions: These results provide adjunctive criteria to evaluate more accurately the fate of a spine instrumentation. Particular care should be placed in the screw evaluation regarding the craniocaudal positioning and alignment.

Download full-text PDF

Source
http://dx.doi.org/10.1016/j.spinee.2013.03.046DOI Listing

Publication Analysis

Top Keywords

pedicle screws
12
screws implanted
12
cortical violation
12
primary stability
8
positioning alignment
8
axial pullout
8
pullout forces
8
screws
6
screw
5
cortical
5

Similar Publications

Want AI Summaries of new PubMed Abstracts delivered to your In-box?

Enter search terms and have AI summaries delivered each week - change queries or unsubscribe any time!