Background: Mechanical torque devices (MTDs) are one of the most commonly recommended devices used to deliver optimal torque to the screw of dental implants. Recently, high variability has been reported about the accuracy of spring-style mechanical torque devices (S-S MTDs). Joint stability and survival rate of fixed implant supported prosthesis depends on the accuracy of these devices. Currently, there is limited information on the steam sterilization influence on the accuracy of MTDs. The purpose of this study was to assess the effect of steam sterilization on the accuracy (±10% of the target torque) of spring-style mechanical torque devices for dental implants.

Materials And Methods: Fifteen new S-S MTDs and their appropriate drivers from three different manufacturers (Nobel Biocare, Straumann [ITI], and Biomet 3i [3i]) were selected. Peak torque of devices (5 in each subgroup) was measured before and after autoclaving using a Tohnichi torque gauge. Descriptive statistical analysis was used and a repeated-measures ANOVA with type of device as a between-subject comparison was performed to assess the difference in accuracy among the three groups of spring-style mechanical torque devices after sterilization. A Bonferroni post hoc test was used to assess pairwise comparisons.

Results: Before steam sterilization, all the tested devices stayed within 10% of their target values. After 100 sterilization cycles, results didn't show any significant difference between raw and absolute error values in the Nobel Biocare and ITI devices; however the results demonstrated an increase of error values in the 3i group (P < 0.05). Raw error values increased with a predictable pattern in 3i devices and showed more than a 10% difference from target torque values (maximum difference of 14% from target torque was seen in 17% of peak torque measurements).

Conclusion: Within the limitation of this study, steam sterilization did not affect the accuracy (±10% of the target torque) of the Nobel Biocare and ITI MTDs. Raw error values increased with a predictable pattern in 3i devices and showed more than 10% difference from target torque values. Before expanding upon the clinical implications, the controlled and combined effect of aging (frequency of use) and steam sterilization needs more investigation.

Download full-text PDF

Source
http://www.ncbi.nlm.nih.gov/pmc/articles/PMC3652367PMC
http://dx.doi.org/10.2147/CCIDEN.S32052DOI Listing

Publication Analysis

Top Keywords

steam sterilization
24
torque devices
24
mechanical torque
20
target torque
20
spring-style mechanical
16
error values
16
torque
14
devices
12
nobel biocare
12
sterilization accuracy
8

Similar Publications

The Influence of the Steam Sterilization Process on Selected Properties of Polymer Samples Produced in MEX and JMT Processes.

Materials (Basel)

November 2024

Institute of Robots and Machines Design, Faculty of Mechanical Engineering, Military University of Technology, gen. Sylwestra Kaliskiego 2 St., 00-908 Warsaw, Poland.

Polymeric materials are widely used in medical engineering, and with the dynamic development of additive manufacturing (AM) technology, increasing attention is being paid to research on the mechanical strength of composite polymer structures. At the same time, the impact of sterilization on, for example, surgical templates and the influence of the sterilization process on the geometry of these parts have not been sufficiently studied. In this work, the effect of steam sterilization on samples made of polymer materials for medical applications was presented.

View Article and Find Full Text PDF

Psychological Effects of Noise Exposure on Personnel in Central Sterile Supply Department in China.

Psychol Res Behav Manag

December 2024

Department of Central Sterile Supply, Hainan General Hospital, Hainan Affiliated Hospital of Hainan Medical University, Haikou, 570311, People's Republic of China.

Objective: This study aimed to investigate whether noise exposure in hospital sterile supply centers exacerbated the psychological burden on staff, potentially mediated by concerns about physiological health.

Methods: Between July 5 and August 4, 2023, 308 staffs from central sterile supply departments in 30 hospitals in Hainan were chosen by convenient sampling method. They completed self-administered questionnaires to assess noise exposure status and used the Self-Reporting Inventory-90 to quantify the psychological state of staff across three dimensions: somatization, anxiety, and sleep and diet.

View Article and Find Full Text PDF

The presence of Bacillus cereus in spices and herbs has posed a detrimental effect on food safety. The absence of thorough testing, comprehensive reporting, and vigilant surveillance of the illness has resulted in a significant underestimation of the true prevalence of foodborne illness caused by B. cereus.

View Article and Find Full Text PDF

Polysaccharide-based hydrogels have attracted significant attention in fields such as drug delivery, tissue engineering, and wound healing, primarily due to their excellent water-retention capacity, biocompatibility, and biodegradability. This study reports the preparation of a hydrogel through the copolymerization of acrylamide (AM) and carboxymethyl cellulose (CMC) using a simple free-radical polymerization method. β-cyclodextrin (β-CD), capable of encapsulating hydrophobic drugs, was chemically modified with double bonds (Ac-β-CD) and incorporated as a pendant unit in the polymerization reaction, forming a ternary copolymer hydrogel with CMC, AM, and Ac-β-CD.

View Article and Find Full Text PDF

Pharmaceutical preclinical tests using cell cultures are nowadays commonly automated. Incubator microbial contaminations impact such tests. Chlorine dioxide (ClO) is widely used in aqueous solutions.

View Article and Find Full Text PDF

Want AI Summaries of new PubMed Abstracts delivered to your In-box?

Enter search terms and have AI summaries delivered each week - change queries or unsubscribe any time!