Impact of wave front and coherence optimization in coherent diffractive imaging.

Opt Express

Commissariat à l’Energie Atomique, Service des Photons, Atomes et Molécules, Bâtiment 522, Centre d’Etude de Saclay, 91191 Gif-sur-Yvette, France.

Published: May 2013

We present single shot nanoscale imaging using a table-top femtosecond soft X-ray laser harmonic source at a wavelength of 32 nm. We show that the phase retrieval process in coherent diffractive imaging critically depends on beam quality. Coherence and image fidelity are measured from single-shot coherent diffraction patterns of isolated nano-patterned slits. Impact of flux, wave front and coherence of the soft X-ray beam on the phase retrieval process and the image quality are discussed. After beam improvements, a final image reconstruction is presented with a spatial resolution of 78 nm (half period) in a single 20 fs laser harmonic shot.

Download full-text PDF

Source
http://dx.doi.org/10.1364/OE.21.011441DOI Listing