In-house developed ELISA for HbA1c was evaluated in comparison with Bio-Rad D-10 Dual Program (HPLC). % HbA1c was estimated using said methods in 40 blood samples. Intra and inter assay coefficient of variation (CV) were in acceptable range i.e., <3%. HbA1c results obtained by two methods showed good correlation, (y = 0.8942x + 0.7229 R2 = 0.877). Sensitivity, specificity, positive, and negative predictive value for in-house ELISA classifying patients as controlled and uncontrolled diabetes against HPLC were recorded as 87%, 94.1%, 95.0%, and 93.8%, respectively, while accuracy of the test was 90.0%. In-house ELISA for HbA1c is not only accurate, cost effective and simple but also precise comparable to HPLC in deciding glycemic status of patients.