In this work, we adapted a semi-automatic segmentation algorithm for vascular structures to extract cerebral blood vessels in the 3D intraoperative contrast-enhanced ultrasound angiographic (3D-iUSA) data of the brain. We quantitatively evaluated the segmentation method with a physical vascular phantom. The geometrical features of the segmentation model generated by the algorithm were compared with the theoretical tube values and manual delineations provided by observers. For a silicon tube with a radius of 2 mm, the results showed that the algorithm overestimated the lumen radii values by about 1 mm, representing one voxel in the 3D-iUSA data. However, the observers were more hindered by noise and artifacts in the data, resulting in a larger overestimation of the tube lumen (twice the reference size). The first results on 3D-iUSA patient data showed that the algorithm could correctly restitute the main vascular segments with realistic geometrical features data, despite noise, artifacts and unclear blood vessel borders. A future aim of this work is to provide neurosurgeons with a visualization tool to navigate through the brain during aneurysm clipping operations.

Download full-text PDF

Source
http://dx.doi.org/10.1515/bmt-2012-0089DOI Listing

Publication Analysis

Top Keywords

semi-automatic segmentation
8
segmentation algorithm
8
3d-iusa data
8
geometrical features
8
noise artifacts
8
algorithm
5
data
5
evaluation semi-automatic
4
segmentation
4
algorithm intraoperative
4

Similar Publications

Want AI Summaries of new PubMed Abstracts delivered to your In-box?

Enter search terms and have AI summaries delivered each week - change queries or unsubscribe any time!