Anterior internal fixator versus a femoral distractor and external fixation for sacroiliac joint compression and single stance gait testing: a mechanical study in synthetic bone.

Int Orthop

Orthopaedic Surgery, Detroit Medical Center, 4D4 University Health Center, Detroit Receiving Hospital, Wayne State University, 4201 St. Antoine Blvd., Detroit, MI 48201, USA.

Published: July 2013

Purpose: The purpose of this study was to evaluate the biomechanical stability and compressive forces across the sacroiliac (SI) joint of an anterior internal fixator compared to the femoral distractor and external fixator for vertically unstable pelvic fractures.

Methods: Five composite pelvises with a simulated APC type III injury fixed with a femoral distractor, external fixator, or anterior internal fixator were tested. A pressure-sensitive film (Tekscan) was placed in the disrupted SI joint recording the magnitude of force. Then, in a single-leg stance model (Instron machine), a load was applied through the sacrum. We recorded displacement at the pubic symphysis and SI joint using high-speed video. Peak load and displacement were measured, and axial stiffness was calculated. Values were compared using a Student's t-test (p < 0.05).

Results: The SI joint was compressed significantly (p < 0.001) more using the anterior internal fixator (18.9 N) and femoral distractor (18.6 N) than the two-pin external fixator (2.5 N). There was no significant difference between the anterior internal fixator and the femoral distractor in displacement at the SI joint. The pubic symphysis displaced less with the femoral distractor than the anterior internal fixator (5.5 mm vs. 4.1 mm; p < 0.05).

Conclusions: The anterior pedicle screw internal fixator allows for indirect compression across the sacroiliac joint that is superior to two-pin external fixation and comparable to the femoral distractor. The anterior internal fixator may be an option for temporary anterior pelvic fixation in situations where external fixation or the femoral distractor have otherwise been used.

Download full-text PDF

Source
http://www.ncbi.nlm.nih.gov/pmc/articles/PMC3685661PMC
http://dx.doi.org/10.1007/s00264-013-1890-9DOI Listing

Publication Analysis

Top Keywords

internal fixator
32
femoral distractor
32
anterior internal
28
distractor external
12
external fixation
12
sacroiliac joint
12
external fixator
12
fixator
11
anterior
9
femoral
8

Similar Publications

Want AI Summaries of new PubMed Abstracts delivered to your In-box?

Enter search terms and have AI summaries delivered each week - change queries or unsubscribe any time!