Notching of the femoral stem neck in metal-on-metal total hip replacement: a case report.

J Orthop Surg (Hong Kong)

Department of Orthopaedics and Traumatology, Queen Mary Hospital, Hong Kong.

Published: April 2013

We report a case of impingement in a metal-on-metal total hip replacement causing both notching of the femoral stem neck and aseptic loosening of the acetabular component. The acetabular component was inserted in excessive anteversion. The femoral stem and acetabular components were retrieved. A larger femoral head was used, and an acetabular component was placed in a less anteverted position. Intra-operative testing through different ranges of movement is recommended to identify potential impingement.

Download full-text PDF

Source
http://dx.doi.org/10.1177/230949901302100129DOI Listing

Publication Analysis

Top Keywords

femoral stem
12
acetabular component
12
notching femoral
8
stem neck
8
metal-on-metal total
8
total hip
8
hip replacement
8
neck metal-on-metal
4
replacement case
4
case report
4

Similar Publications

Uncontrollable non-compressible hemorrhage and traumatic infection have been major causes of mortality and disability in both civilian and military populations. A dressing designed for point-of-care control of non-compressible hemorrhage and prevention of traumatic infections represents an urgent medical need. Here, a novel self-gelling sponge OHN@ε-pL is developed, integrating N-succinimidyl ester oxidized hyaluronic acid (OHN) and ε-poly-L-lysine (ε-pL).

View Article and Find Full Text PDF

Short-stem total hip arthroplasty (SHA) has become popular because it preserves femoral bone stock and enables the use of short femoral stems in revision total hip arthroplasty (THA). However, no study has evaluated whether a short stem in revision THA, replacing a standard stem, can provide adequate primary stability to facilitate osseous integration. In this biomechanical study, a metaphyseal anchoring SHA (Tri-Lock BPS) stem and a standard THA (Corail) stem were implanted into ten composite femurs and loaded dynamically from 300 to 1700 N with 1 Hz.

View Article and Find Full Text PDF

Osteoporosis is caused by an imbalance between bone resorption and formation, which decreases bone mass and strength and increases the risk of fracture. Therefore, osteoporosis is treated with oral resorption inhibitors, such as bisphosphonates, and parenteral osteogenic drugs, including parathyroid hormone and antisclerostin antibodies. However, orally active osteogenic drugs have not yet been developed.

View Article and Find Full Text PDF

Background: Improvements in cementless total hip arthroplasty have been directed at optimizing osseointegration of the femoral implant to reduce aseptic loosening rates. Stem design plays a critical role in the performance of these implants. Given the increase in new stem designs and the creation of an updated classification system, improved understanding of the outcomes of each stem type is warranted.

View Article and Find Full Text PDF

Proximal femoral fractures are particularly common in older adults, and cases requiring conversion to total hip arthroplasty may arise because of treatment failure or osteoarthritis. Fractures around the distal screw removal holes can be problematic. This study aimed to analyze the relationship between stem length and femoral stress distribution to determine the optimal stem length.

View Article and Find Full Text PDF