We evaluated the ability of four commercial MIC testing systems (MicroScan, Vitek 2, Phoenix, and Etest) to detect vancomycin MIC values of ≤1 to ≥2 in 200 methicillin-resistant Staphylococcus aureus (MRSA) strains compared to the Clinical and Laboratory Standards Institute broth microdilution (BMD) reference methods. Compared to the BMD method, absolute agreement (0 ± dilution) was highest for the Phoenix system (66.2%) and the MicroScan turbidity method (61.8%), followed by the Vitek 2 system (54.3%). The Etest produced MIC values 1 to 2 dilutions higher than those produced by the BMD method (36.7% agreement). Of interest, the MicroScan system (prompt method) was more likely to overcall an MIC value of 1 mg/liter (74.1%), whereas the Phoenix (76%) and Vitek 2 (20%) systems had a tendency to undercall an MIC of 2 mg/liter. The ability to correctly identify vancomycin MIC values of 1 and 2 has clinical implications and requires further evaluation.

Download full-text PDF

Source
http://www.ncbi.nlm.nih.gov/pmc/articles/PMC3697692PMC
http://dx.doi.org/10.1128/JCM.00448-13DOI Listing

Publication Analysis

Top Keywords

mic values
12
methicillin-resistant staphylococcus
8
staphylococcus aureus
8
vancomycin mic
8
bmd method
8
mic mg/liter
8
mic
6
evaluation vancomycin
4
vancomycin susceptibility
4
susceptibility testing
4

Similar Publications

Want AI Summaries of new PubMed Abstracts delivered to your In-box?

Enter search terms and have AI summaries delivered each week - change queries or unsubscribe any time!