Purpose: Iterative reconstruction has recently been revisited as a promising concept for substantial CT dose reduction. The purpose of this study was to assess the potential benefit of sinogram affirmed iterative reconstruction (SAFIRE) in head CT by comparing objective and subjective image quality at reduced tube current with standard dose filtered back projection (FBP).
Materials And Methods: Non-contrast reduced dose head CT (255 mAs, CTDIvol 47.8 mGy) was performed in thirty consecutive patients and reconstructed with SAFIRE and FBP. Images were assessed in terms of quantitative and qualitative image quality and compared with FBP of standard dose acquisitions (320 mAs, CTDI vol 59.7 mGy).
Results: In reduced dose CT examinations, use of SAFIRE versus FBP resulted in 47% increase in contrast-to-noise ratio (CNR) (2.49 vs. 1.69; p<0.0001). While reduction of tube current was associated with 13% decrease in CNR, quantitative degradation of image quality at lower dose was more than compensated through SAFIRE (2.49 vs. 1.96; p=0.0004). Objective measurements of image sharpness were comparable between FBP and SAFIRE reconstructions (575.9 ± 74.1 vs. 583.4 ± 74.7 change in HU/Pixel; p=0.28). Compared to standard dose FBP, subjective grading of noise as well as overall image quality scores were significantly improved when SAFIRE was used in reduced dose exams (1.3 vs. 1.6, p=0.006; 1.3 vs. 1.7, p=0.026).
Conclusion: At 20% dose reduction, reconstruction of head CT by SAFIRE provides above standard objective and subjective image quality, suggesting potential for more vigorous dose savings in neuroradiology CT applications.
Download full-text PDF |
Source |
---|---|
http://dx.doi.org/10.1016/j.ejrad.2013.03.011 | DOI Listing |
Photoacoustics
February 2025
School of Biomedical Engineering, Southern Medical University, Guangzhou, Guangdong 510515, China.
Photoacoustic tomography (PAT) enables non-invasive cross-sectional imaging of biological tissues, but it fails to map the spatial variation of speed-of-sound (SOS) within tissues. While SOS is intimately linked to density and elastic modulus of tissues, the imaging of SOS distribution serves as a complementary imaging modality to PAT. Moreover, an accurate SOS map can be leveraged to correct for PAT image degradation arising from acoustic heterogeneities.
View Article and Find Full Text PDFJpn J Radiol
January 2025
Department of Radiology, Ehime University Graduate School of Medicine, Shitsukawa, Toon, Ehime, 791-0295, Japan.
Purpose: To evaluate the effects of four-dimensional noise reduction filtering using a similarity algorithm (4D-SF) on the image quality and tumor visibility of low-dose dynamic computed tomography (CT) in evaluating breast cancer.
Materials And Methods: Thirty-four patients with 38 lesions who underwent low-dose dynamic breast CT and were pathologically diagnosed with breast cancer were enrolled. Dynamic CT images were reconstructed using iterative reconstruction alone or in combination with 4D-SF.
Phys Med
January 2025
Medical Physics Unit, ASST Monza, Monza, Italy.
Purpose: Digital Breast Tomosynthesis (DBT) is an advanced mammography technique for which there are currently no internationally agreed methods and reference values for image quality assessment. The aim of this multicentre study was to evaluate a simple method to assess the technical image quality of reconstructed and synthetic 2D (SM) images of different models of DBT systems using commercially available phantoms.
Methods: The signal difference to noise ratio (SDNR) was chosen as an index of technical image quality and was evaluated for three commercial phantoms, Tomophan, Tormam and CIRS model 015, on 55 DBT systems (six vendors, nine models).
Quant Imaging Med Surg
January 2025
Department of Imaging Medicine and Nuclear Medicine, Shandong Second Medical University, Weifang, China.
Background: Rapid kilovolt (kV)-switching dual-energy computed tomography (DECT) has been increasingly applied to the measurement of lumbar spine bone mineral density (BMD) in humans and animal models. The objective of this study was to investigate the optimal parameters for the measurement of vertebral BMD. The BMD of the spinal model was measured by means of DECT in combination with different noise index (NI) and preset adaptive statistical iterative reconstruction Veo (ASiR-V) levels.
View Article and Find Full Text PDFQuant Imaging Med Surg
January 2025
Henan Key Laboratory of Imaging and Intelligent Processing, Information Engineering University, Zhengzhou, China.
Background: Photon-counting computed tomography (CT) is an advanced imaging technique that enables multi-energy imaging from a single scan. However, the limited photon count assigned to narrow energy bins leads to increased quantum noise in the reconstructed spectral images. To address this issue, leveraging the prior information in the spectral images is essential.