Background: Posterior cruciate ligament (PCL) retaining (CR) and -sacrificing (PS) total knee arthroplasties (TKA) are widely-used to treat osteoarthritis of the knee joint. The PS design substitutes the function of the PCL with a cam-spine mechanism which may produce adverse changes to joint kinematics and kinetics.

Methods: CR- and PS-TKA were performed on 11 human knee specimens. Joint kinematics were measured with a dynamic knee simulator and motion tracking equipment. In-situ loads of the PCL and cam-spine were measured with a robotic force sensor system. Partial weight bearing flexions were simulated and external forces were applied.

Results: The PS-TKA rotated significantly less throughout the whole flexion range compared to the CR-TKA. Femoral roll back was greater in the PS-TKA; however, this was not correlated with lower quadriceps forces. Application of external loads produced significantly different in-situ force profiles between the TKA systems.

Conclusions: Our data demonstrate that the PS-design significantly alters kinematics of the knee joint. Our data also suggest the cam-spine mechanism may have little influence on high flexion kinematics (such as femoral rollback) with most of the load burden shared by supporting implant and soft-tissue structures.

Download full-text PDF

Source
http://dx.doi.org/10.1016/j.knee.2013.03.005DOI Listing

Publication Analysis

Top Keywords

knee joint
12
joint kinematics
12
posterior cruciate
8
total knee
8
pcl cam-spine
8
cam-spine mechanism
8
knee
6
joint
5
kinematics
5
differences knee
4

Similar Publications

Want AI Summaries of new PubMed Abstracts delivered to your In-box?

Enter search terms and have AI summaries delivered each week - change queries or unsubscribe any time!