Aim: To compare the international normalised ratio (INR) value of patients evaluated using the CoaguChek XS versus conventional laboratory methods, in the period after open-heart surgery for mechanical valve replacement until a therapeutic range is achieved using vitamin K antagonists (VKA) together with low molecular weight heparin (LMWH).
Methods: One hundred and five patients undergoing open-heart surgery for mechanical valve replacement were enrolled. Blood samples were collected from patients before surgery, and on the second and fifth postoperative days, simultaneously for both the point of care device and conventional laboratory techniques. Patients were administered VKA together with LMWH at therapeutic doses (enoxaparin 100 IU/kg twice daily) subcutaneously, until an effective range was achieved on approximately the fifth day after surgery.
Results: The mean INR values using the CoaguChek XS preoperatively and on the second and fifth days postoperatively were 1.20 (SD ± 0.09), 1.82 (SD ± 0.45), and 2.55 (SD ± 0.55), respectively. Corresponding results obtained using conventional laboratory techniques were 1.18 (SD ± 0.1), 1.81 (SD ± 0.43), and 2.51 (SD ± 0.58). The correlation coefficient was r = 0.77 preoperatively, r = 0.981 on postoperative day 2, and r = 0.983 on postoperative day 5.
Discussion: Results using the CoaguChek XS Handheld Coagulation Analyzer correlated strongly with conventional laboratory methods, in the bridging period between open-heart surgery for mechanical valve replacement and the achievement of a therapeutic range on warfarin and LMWH.
Download full-text PDF |
Source |
---|---|
http://dx.doi.org/10.1111/jocs.12100 | DOI Listing |
Enter search terms and have AI summaries delivered each week - change queries or unsubscribe any time!