Analyses of the results of calibration of survey meters carried out at the Secondary Standards Dosimetry Laboratory (SSDL) in Ghana over a period of 4 y (2008-2011) are reported. The calibration factors (CFs) of the set of survey meters indicated that ∼91.04 % were within the acceptable limit of ± 20.0 %. A higher percentage of the survey meters indicated CFs in the range of 0.95-1.15 except a few of them which indicated values <0.55. Some of the survey meters also recorded CFs >1.15. The degree of uncertainty in the measurements ranged from 0.03 to 17 % with the majority of them ranging from 0.03 to 6.0 % and a few of them >6.0 %. The results show that most of the survey meters calibrated were within the requirements of the regulations and may provide data for future development of calibration techniques in the country.