Purpose: A dedicated stereotactic gamma irradiation device, the GammaPod™ from Xcision Medical Systems, was developed specifically to treat small breast cancers. This study presents the first evaluation of dosimetric and geometric characteristics from the initial prototype installed at University of Maryland Radiation Oncology Department.

Methods: The GammaPod™ stereotactic radiotherapy device is an assembly of a hemi-spherical source carrier containing 36 (60)Co sources, a tungsten collimator, a dynamically controlled patient support table, and the breast immobilization system which also functions as a stereotactic frame. The source carrier contains the sources in six columns spaced longitudinally at 60° intervals and it rotates together with the variable-size collimator to form 36 noncoplanar, concentric arcs focused at the isocenter. The patient support table enables motion in three dimensions to position the patient tumor at the focal point of the irradiation. The table moves continuously in three cardinal dimensions during treatment to provide dynamic shaping of the dose distribution. The breast is immobilized using a breast cup applying a small negative pressure, where the immobilization cup is embedded with fiducials also functioning as the stereotactic frame for the breast. Geometric and dosimetric evaluations of the system as well as a protocol for absorbed dose calibration are provided. Dosimetric verifications of dynamically delivered patient plans are performed for seven patients using radiochromic films in hypothetical preop, postop, and target-in-target treatment scenarios.

Results: Loaded with 36 (60)Co sources with cumulative activity of 4320 Ci, the prototype GammaPod™ unit delivers 5.31 Gy/min at the isocenter using the largest 2.5 cm diameter collimator. Due to the noncoplanar beam arrangement and dynamic dose shaping features, the GammaPod™ device is found to deliver uniform doses to targets with good conformity. The spatial accuracy of the device to locate the radiation isocenter is determined to be less than 1 mm. Single shot profiles with 2.5 cm collimator are measured with radiochromic film and found to be in good agreement with respect to the Monte Carlo based calculations (congruence of FWHM less than 1 mm). Dosimetric verifications corresponding to all hypothetical treatment plans corresponding to three target scenarios for each of the seven patients demonstrated good agreement with gamma index pass rates of better than 97% (99.0% ± 0.7%).

Conclusions: Dosimetric evaluation of the first GammaPod™ stereotactic breast radiotherapy unit was performed and the dosimetric and spatial accuracy of this novel technology is found to be feasible with respect to clinical radiotherapy standards. The observed level of agreement between the treatment planning system calculations and dosimetric measurements has confirmed that the system can deliver highly complex treatment plans with remarkable geometric and dosimetric accuracy.

Download full-text PDF

Source
http://dx.doi.org/10.1118/1.4794477DOI Listing

Publication Analysis

Top Keywords

dosimetric
9
dosimetric geometric
8
stereotactic radiotherapy
8
radiotherapy device
8
gammapod™ stereotactic
8
source carrier
8
60co sources
8
patient support
8
support table
8
stereotactic frame
8

Similar Publications

Technical Aspects of a National Service for Paediatric Pelvic Brachytherapy.

Clin Oncol (R Coll Radiol)

December 2024

Department of Oncology, University College London Hospitals NHS Foundation Trust, 250 Euston Road, London, NW1 2PG, UK. Electronic address:

Aims: Brachytherapy is advantageous for localised rhabdomyosarcomas in children compared with external beam radiotherapy, sparing close organs at risk with highly conformal dosimetry. A methodology for planning and delivering fractionated high-dose-rate paediatric pelvic brachytherapy is detailed, and the dosimetric parameters are presented. This provides a practical template for radiotherapy departments with a similar patient cohort to implement this treatment technique.

View Article and Find Full Text PDF

Aim: To study the dosimetric behavior of dose computational algorithms in inhomogeneous medium using CMS XiO and MONACO treatment planning system (TPS) for 4 megavoltage (MV), 6 MV and 15 MV photon beam energies.

Material And Methods: Styrofoam blocks of thickness 1.90 cm, 3.

View Article and Find Full Text PDF

Dosimetric comparison of VMAT plans in preoperative short-course rectal radiotherapy.

Sci Rep

December 2024

Sir Run Run Shaw Hospital, School of Medicine, Zhejiang University, Hangzhou, 310000, China.

Recently, neoadjuvant short-course radiation therapy (SCRT) has emerged as a valid treatment option for patients with locally advanced rectal cancer (LARC). We assessed SCRT plans using volumetric-modulated arc therapy (VMAT) with Halcyon and Infinity medical linear accelerators (Linacs) and compared the plan quality and delivery efficiency across all cases. Thirty patients who underwent preoperative SCRT for LARC at the hospital were randomly selected.

View Article and Find Full Text PDF

Development of a prediction model for tube feeding dependence in HPV-associated oropharyngeal cancer patients undergoing chemoradiotherapy.

Oral Oncol

December 2024

Radiation Medicine Program, Princess Margaret Cancer Centre, M5G 2M9, Toronto, Ontario, Canada; Department of Radiation Oncology, Princess Margaret Cancer Centre, University of Toronto, M5G 2M9 Toronto, Ontario. Electronic address:

Objectives: This study aimed to develop a prediction model for feeding tube dependence in a large homogenous cohort of HPV-associated oropharyngeal squamous cell carcinoma (HPV + OPSCC) patients receiving chemoradiotherapy (CRT). We further aimed to externally validate three previously published feeding tube prediction models on this cohort.

Materials And Methods: p16-confirmed HPV + OPSCC patients treated with definitive CRT at a tertiary cancer centre between April 2017 and February 2022 were identified.

View Article and Find Full Text PDF

. Adult medulloblastoma (AMB) patients should receive postoperative craniospinal irradiation (CSI) as a standard treatment. Volumetric intensity-modulated arc therapy (VMAT) is a promising method for CSI.

View Article and Find Full Text PDF

Want AI Summaries of new PubMed Abstracts delivered to your In-box?

Enter search terms and have AI summaries delivered each week - change queries or unsubscribe any time!