In vivo evaluation of the iPex and Root ZX electronic apex locators using various irrigants.

Int Endod J

Department of Restorative Dentistry and Endodontics, Universitat Internacional de Catalunya, Sant Cugat del Valles, Barcelona, Spain.

Published: August 2013

Aim: To evaluate in vivo the performance of the iPex and Root ZX electronic apex locators (EALs) in the presence of several irrigant solutions: 2.5% sodium hypochlorite (NaOCl) and 2% chlorhexidine (CHX).

Methodology: Thirty-two single-rooted human teeth that were scheduled for extraction were selected. Teeth with metallic restorations, fractures, root resorption, pulp necrosis or open apices were not included The working length (WL) was determined electronically for the root canals with the iPex and Root ZX EALs in the presence of two different irrigant solutions, 2.5% NaOCl and 2% CHX. After the teeth had been extracted, a size 10 K-file was used to determine the reference working length (RWL), which was established at 0.5 mm short of the major foramen. In each case, the RWL was subtracted from the electronic measurements. Positive values indicated electronic measurements that exceeded the RWL (long measurements), whereas negative values indicated measurements that were short of the RWL. The values obtained with the different irrigants and EALs were compared using the paired t-test. Significance was set at P < 0.05.

Results: The accuracy of the iPex nor Root ZX EAL was not affected by 2.5% NaOCl or 2% CHX (P > 0.05). However, significant differences were observed between the readings of the iPex and Root ZX, irrespective of whether 2.5% NaOCl or 2% CHX was used as the irrigant (P < 0.05). The iPex was less accurate than the Root ZX in determining the RWL.

Conclusions: The accuracy of neither the iPex nor Root ZX EAL was affected by the irrigant used. However, the iPex was less accurate than the Root ZX in determining the RWL both for 2.5% NaOCl and for 2% CHX.

Download full-text PDF

Source
http://dx.doi.org/10.1111/iej.12057DOI Listing

Publication Analysis

Top Keywords

ipex root
16
root electronic
8
electronic apex
8
apex locators
8
eals presence
8
presence irrigant
8
irrigant solutions
8
solutions 25%
8
working length
8
25% naocl
8

Similar Publications

Aim: An accurate working length (WL) estimation is fundamental to a successful endodontic therapy. The objective of this comparative research was to investigate the performance of iPex II and the Root ZX Mini electronic apex locators (EALs) in measuring root canal WL with different irrigant solutions and their relation to the electrical conductivity of irrigation solutions.

Materials And Methods: Seventy sound permanent lower premolar teeth, each with a single root and developed apices, were used.

View Article and Find Full Text PDF

Influence of Embedding Medium on iPex Electronic Apex Locator Accuracy: An Comparative Study Using Alginate, Saline, and Gelatin Models.

J Contemp Dent Pract

October 2023

Department of Endodontics, Faculty of Dentistry, King Abdulaziz University, Jeddah, Saudi Arabia, Phone: +966555531081, e-mail: https://orcid.org/0000-0003-0817-7129.

Aim: Accurate working length determination during root canal treatment is essential for achieving successful outcomes. This study aimed to evaluate the impact of embedding medium on the accuracy of iPex electronic apex locator (EAL).

Materials And Methods: Sixty-one extracted single-rooted teeth were decoronated and coronally flared with Gates-Glidden burs.

View Article and Find Full Text PDF

Objective: To investigate the effect of the hand file size on the accuracy of Root ZX (J. Morita Co., Kyoto, Japan) and iPex (NSK, Tochigi, Japan).

View Article and Find Full Text PDF

Evaluation of the accuracy of different apex locators in determining the working length during root canal retreatment.

J Dent Res Dent Clin Dent Prospects

June 2020

Department of Endodontics, Faculty of Dentistry, Ankara University, Ankara, Turkey.

This study aimed to assess the accuracy of three electronic apex locators (EALs) (Dentaport ZX [J Morita, Tokyo, Japan], Propex Pixi [Dentsply Maillefer, Ballaigues, Switzerland], and iPex II [NSK, Tokyo, Japan]) during root canal retreatment. The root canal lengths of 90 extracted single-rooted human teeth were determined under a dental operating microscope at ×10 magnification. The actual working length (AWL) was 0.

View Article and Find Full Text PDF

Electronic apex locators (EALs), cone-beam computed tomography (CBCT), and radiovisiography (RVG) claim to determine of working length (WL) in primary molar teeth. To evaluate the accuracy of WL determination by using three EALs, RVG, and CBCT imaging in extracted human primary molar teeth with/without root resorption. Twenty extracted primary molar teeth with resorption and 20 extracted primary molar teeth without resorption were used.

View Article and Find Full Text PDF

Want AI Summaries of new PubMed Abstracts delivered to your In-box?

Enter search terms and have AI summaries delivered each week - change queries or unsubscribe any time!