A safe zone for acetabular component positioning in hip resurfacing (RAIL: Relative Acetabular Inclination Limit) was calculated based on implant size and acetabular inclination angle (AIA). For AIA below the RAIL, there were no adverse wear failures or dislocations, and only 1% of cases with ion levels above 10 μg/L. Other than high inclination angle and small bearing size, female gender was the only other factor that correlated with high ion levels in the multivariate analysis. Seven hundred sixty-one hip resurfacing cases are included in this study. The UCLA activity score, femoral shaft angle, body mass index, weight, American Society of Anesthesiologists score, combined range of motion, diagnosis, age, gender, implant brand, AIA, bearing size, and duration of implantation were analyzed to determine the potential risk factors for elevated metal ion levels. These findings apply to sub hemispheric metal-on-metal bearings with similar coverage arcs as the Biomet and Corin hip resurfacing brands. Additional problems may occur when these bearings are connected with trunions on stems for total hip arthroplasty.