Purpose: The authors have recently developed a high-resolution microscope-integrated spectral domain optical coherence tomography (MIOCT) device designed to enable OCT acquisition simultaneous with surgical maneuvers. The purpose of this report is to describe translation of this device from preclinical testing into human intraoperative imaging.

Methods: Before human imaging, surgical conditions were fully simulated for extensive preclinical MIOCT evaluation in a custom model eye system. Microscope-integrated spectral domain OCT images were then acquired in normal human volunteers and during vitreoretinal surgery in patients who consented to participate in a prospective institutional review board-approved study. Microscope-integrated spectral domain OCT images were obtained before and at pauses in surgical maneuvers and were compared based on predetermined diagnostic criteria to images obtained with a high-resolution spectral domain research handheld OCT system (HHOCT; Bioptigen, Inc) at the same time point. Cohorts of five consecutive patients were imaged. Successful end points were predefined, including ≥80% correlation in identification of pathology between MIOCT and HHOCT in ≥80% of the patients.

Results: Microscope-integrated spectral domain OCT was favorably evaluated by study surgeons and scrub nurses, all of whom responded that they would consider participating in human intraoperative imaging trials. The preclinical evaluation identified significant improvements that were made before MIOCT use during human surgery. The MIOCT transition into clinical human research was smooth. Microscope-integrated spectral domain OCT imaging in normal human volunteers demonstrated high resolution comparable to tabletop scanners. In the operating room, after an initial learning curve, surgeons successfully acquired human macular MIOCT images before and after surgical maneuvers. Microscope-integrated spectral domain OCT imaging confirmed preoperative diagnoses, such as full-thickness macular hole and vitreomacular traction, and demonstrated postsurgical changes in retinal morphology. Two cohorts of five patients were imaged. In the second cohort, the predefined end points were exceeded with ≥80% correlation between microscope-mounted OCT and HHOCT imaging in 100% of the patients.

Conclusion: This report describes high-resolution MIOCT imaging using the prototype device in human eyes during vitreoretinal surgery, with successful achievement of predefined end points for imaging. Further refinements and investigations will be directed toward fully integrating MIOCT with vitreoretinal and other ocular surgery to image surgical maneuvers in real time.

Download full-text PDF

Source
http://www.ncbi.nlm.nih.gov/pmc/articles/PMC3696478PMC
http://dx.doi.org/10.1097/IAE.0b013e3182831293DOI Listing

Publication Analysis

Top Keywords

spectral domain
32
microscope-integrated spectral
28
domain oct
20
surgical maneuvers
16
human
10
preclinical evaluation
8
imaging
8
high-resolution microscope-integrated
8
spectral
8
domain
8

Similar Publications

Purpose: To identify the macular retinal layer thickness changes in polyarteritis nodosa (PAN) patients without pathological findings appearing in color fundus photography (CFP), and to investigate the correlations with disease durations.

Methods: A total of 24 PAN patients who had been for 3 years or more and underwent SD-OCT were recruited from the UK Biobank, with exclusions for diabetes, eye disease, or abnormal CFP findings. Only the right eyes were included, with each PAN patient paired one-to-one with a control matched for age, sex, and ethnicity.

View Article and Find Full Text PDF

Purpose: To report the occurrence of AMN (Acute Macular Neuroretinopathy) in a Behçet Disease (BD) patient during an active systemic inflammatory relapse and to describe the SD-OCT features of this entity.

Patients And Methods: Retrospective observational case report of a patient who presented with an AMN during a BD associated ocular inflammation (Saint Pierre Hospital, Brussels, Belgium). Clinical record and imaging, including infrared reflectance image (IR) and spectral domain optical coherence tomography (SD-OCT), were analyzed.

View Article and Find Full Text PDF

Purpose: To compare a novel high-resolution optical coherence tomography (OCT) with improved axial resolution (High-Res OCT) with conventional spectral-domain OCT (SD-OCT) with regard to their capacity to characterize the disorganization of the retinal inner layers (DRIL) in diabetic maculopathy.

Methods: Diabetic patients underwent multimodal retinal imaging (SD-OCT, High-Res OCT, and color fundus photography). Best-corrected visual acuity and diabetes characteristics were recorded.

View Article and Find Full Text PDF

This report describes a patient with polypoidal choroidal vasculopathy (PCV) with fovea-involving retinal pigment epithelium (RPE) tear that showed tissue remodeling with a good visual outcome. Imaging over the patient's clinical course from 2019 was reviewed. A 74-year-old female presented with left submacular hemorrhage and a large multi-lobular pigment epithelial detachment.

View Article and Find Full Text PDF

Purpose: The purpose of this study is to evaluate the impact of foveal bulge presence on visual acuity (VA) in patients with diabetic macular edema (DME) and retinal vein occlusion (RVO).

Methods: Spectral-domain optical coherence tomography (SD-OCT) scans were conducted on 22 DME patients and 20 RVO patients. Ordinary least squares (OLS) regression was employed to analyze the association between VA and the presence of the foveal bulge, as well as factors such as sex, age, central foveal thickness, various line scans of the fovea, and the external limiting membrane (ELM).

View Article and Find Full Text PDF

Want AI Summaries of new PubMed Abstracts delivered to your In-box?

Enter search terms and have AI summaries delivered each week - change queries or unsubscribe any time!