Evaluation of the Bioline Standard Diagnostics SD immunochromatographic norovirus detection kit using fecal specimens from Australian gastroenteritis incidents.

Diagn Microbiol Infect Dis

Victorian Infectious Diseases Reference Laboratory, North Melbourne, Victoria 3051, Australia.

Published: June 2013

Human norovirus is a major cause of both sporadic cases and outbreaks of gastroenteritis and comprises two main genogroups (GI and GII) which, in turn, comprise a variety of genotypes. The current study examined the efficacy of the Bioline SD kit using fecal material from Australian gastroenteritis incidents. At best, the SD kit had a sensitivity of 62%. Freezing and thawing specimens before testing significantly improved sensitivity. The SD kit had a specificity of 98.6%. Genotype analysis (Open Reading Frame 2) indicated the SD kit could detect a range of genotypes and genotype variants including GI.1, GI.3, GI.4, GII.1, GII.3, GII.4 (unclassified), GII.4 (2006b), GII.4 (2009), GII.4 (2012) and GII.6 but the kit failed to detect GI.2 and GII.2 norovirus. The kit did not cross-react with a number of common fecal viruses including astrovirus, sapovirus, rotavirus or adenovirus. The kit was very easy to use and would be valuable in point-of-care testing.

Download full-text PDF

Source
http://dx.doi.org/10.1016/j.diagmicrobio.2013.02.018DOI Listing

Publication Analysis

Top Keywords

kit
8
kit fecal
8
australian gastroenteritis
8
gastroenteritis incidents
8
evaluation bioline
4
bioline standard
4
standard diagnostics
4
diagnostics immunochromatographic
4
immunochromatographic norovirus
4
norovirus detection
4

Similar Publications

Want AI Summaries of new PubMed Abstracts delivered to your In-box?

Enter search terms and have AI summaries delivered each week - change queries or unsubscribe any time!