Distribution of peri-implant stresses with a countertorque device.

Int J Oral Maxillofac Implants

Department of Prosthodontics, School of Dentistry, University of Sao Paulo, Sao Paulo, Brazil.

Published: February 2014

Purpose: To verify the effectiveness of a countertorque device in dental implants in redistributing stress to the bone-implant interface during tightening of the abutment screw.

Materials And Methods: Two prismatic photoelastic samples containing implants were made, one with a 3.75-mm-diameter implant and the other with a 5.0-mm-diameter implant (both implants had an external-hexagon interface) and the respective abutments were attached (CeraOne). The samples were placed in a support and submitted to torques of 10, 20, 32, and 45 Ncm with an electronic torque meter. The torque application was repeated 10 times on each sample (n = 10) with and without a countertorque device. Photoelastic patterns were detected; thus, a photographic register of each test was selected. The fringe patterns were analyzed at discrete points near the implants' external arch.

Results: In both implants analyzed, a stress gradient reduction was observed through the implant with the countertorque device.

Conclusions: The countertorque device used in this study proved to be effective in reducing the stresses generated in the peri-implant bone tissue during torque application.

Download full-text PDF

Source
http://dx.doi.org/10.11607/jomi.2506DOI Listing

Publication Analysis

Top Keywords

countertorque device
16
torque application
8
countertorque
5
distribution peri-implant
4
peri-implant stresses
4
stresses countertorque
4
device
4
device purpose
4
purpose verify
4
verify effectiveness
4

Similar Publications

Type 2 diabetes interferes with bone remodeling mechanisms, requiring studies to reverse this damage, and resveratrol is a polyphenol with rich properties. This study aimed to characterize the long bone morphology and peri-implant biomechanics of normoglycemic and type 2 diabetic animals treated with resveratrol. Thirty-two male Wistar rats were used and divided into normoglycemic and diabetic with or without treatment.

View Article and Find Full Text PDF

A systematic review on removal of osseointegrated implants: un update.

BMC Oral Health

October 2023

Unit of Prosthodontics, Department of Innovative Technologies in Medicine and Dentistry, University "G. d'Annunzio" of Chieti- Pescara, 66100, Chieti, Italy.

Background: Today dental implants represent an effective therapy in case of partial or total edentulism, with an excellent success rate. Despite the results obtained, there may be biological or mechanical complications during the therapy, which lead to the loss of the implant. This systematic review aims to evaluate the current state of the art in the literature on techniques used for the removal of dental implants.

View Article and Find Full Text PDF

Unlabelled: Based on its mechanical properties, PEEK (polyether-ether-ketone) might be useful in restorative procedures. In oral rehabilitation, its viability has been studied mainly for prostheses and dental implants.

Aim: The aim of this study was to evaluate the fit accuracy of dental implant bars made of either PEEK or cobalt-chrome submitted to cycling mechanics.

View Article and Find Full Text PDF

Background: Abutment screw loosening is a common complication of implant supported prostheses, especially for single crowns. In engineering, anaerobic adhesives (AA) are used to provide chemical locking between screw surfaces, but their application in implantology remains unclear.

Purpose: The purpose of this article is to evaluate, in vitro, the effect of AA on counter-torque of abutment screws for cemented prosthesis on dental implants with external hexagon connection (EHC) and conical connection (CC).

View Article and Find Full Text PDF

This study investigated the impact of a modified implant macrogeometry on peri-implant healing and its effect on bone-related molecules in rats. Eighteen rats received one implant in each tibia: the control group received implants with conventional macrogeometry and the test group received implants with modified macrogeometry. After 30 days, the implants were removed for biomechanical analysis and the bone tissue around them was collected for quantifying gene expression of OPN, Runx2, β-catenin, BMP-2, Dkk1, and RANKL/OPG.

View Article and Find Full Text PDF

Want AI Summaries of new PubMed Abstracts delivered to your In-box?

Enter search terms and have AI summaries delivered each week - change queries or unsubscribe any time!