Background: The cuff pressure for optimal airway sealing with first-generation laryngeal mask airway has been shown to be 40 cm H(2)O in children. Currently, there are no data regarding the ideal intracuff pressure for the laryngeal mask airway Supreme (Supreme) in children.

Objectives: To compare the clinical performance of the laryngeal mask airway supreme with the laryngeal mask airway unique in infants and children.

Materials And Methods: One hundred eighty children were assigned to receive either a Supreme or a laryngeal mask airway-U. We hypothesized higher airway leak pressure with the Supreme at both 40 cm H(2)O and 60 cm H(2)O, when compared with the laryngeal mask airway-U. Ease and time of insertion, insertion attempts, fiber optic examination, quality of airway, efficacy of mechanical ventilation, success of gastric tube placement (Supreme), incidence of gastric insufflation, and complications were also assessed.

Results: Airway leak pressure at an intracuff pressure of 60 cm H(2)O for the Supreme was 17.4 (5.2) vs laryngeal mask airway-U at 18.4 (6.6) cm H(2)O and did not differ when compared to an intracuff pressure of 40 cm H(2)O for both devices; Supreme at 17.2 (5) vs laryngeal mask airway-U at 17.7 (6) cm H(2)O. The laryngeal mask airway-U was associated with higher first-attempt success rates. The Supreme was associated with less gastric insufflation than the laryngeal mask airway-U.

Conclusions: Intracuff pressures of 40 cm H(2)O may be sufficient for the Supreme in children, and there may be no added benefit of an intracuff pressure of 60 cm H(2)O, as leak pressures were similar. The Supreme may be preferred over the laryngeal mask airway-U for its lower rates of gastric insufflation and provision for gastric access when mechanical ventilation is utilized.

Download full-text PDF

Source
http://dx.doi.org/10.1111/pan.12145DOI Listing

Publication Analysis

Top Keywords

laryngeal mask
52
mask airway
24
mask airway-u
24
intracuff pressure
16
laryngeal
13
mask
13
gastric insufflation
12
pressure h2o
12
supreme
11
airway
10

Similar Publications

Background: The use of laryngeal masks (LM) has increased widely today, both in anesthesia and in emergency cases. LM are available as reusable and disposable. Although reuse of disposable LM is not recommended, they are reused again after decontamination with different methods in anesthesia units in some countries.

View Article and Find Full Text PDF

Objective: The supraglottic airway device is a viable alternative to tracheal intubation for elective surgery. To conduct a comparative analysis of the advantages and disadvantages associated with use of the Baska mask and I-gel across various dimensions.

Design: A comprehensive search was conducted across PubMed, Embase, Cochrane Library, Web of Science, and other relevant databases to identify randomised controlled trials (RCTs) involving patients who used the Baska mask and I-gel.

View Article and Find Full Text PDF

Background: Several health care networks have fully adopted second-generation supraglottic airway (SGA) i-gel. Real-world evidence of enhanced patient safety after such practice change is lacking. We hypothesized that the implementation of i-gel compared to the previous LMA®-Unique™ would be associated with a lower risk of airway-related safety events.

View Article and Find Full Text PDF

Objective: To evaluate the clinical efficacy and safety of ultrasound-guided rectus sheath block (RSB) in laparoscopic umbilical hernia repair with intraperitoneal onlay mesh (IPOM).

Methods: A total of 139 patients scheduled for laparoscopic umbilical hernia repair with IPOM were selected and randomly assigned to either the group receiving general anesthesia combined with bilateral rectus sheath block (Group GR, 71 patients) or the group receiving general anesthesia alone (Group G, 68 patients). We monitored the patients' heart rate (HR) and mean arterial pressure (MAP) at four time points: before anesthesia induction (T1), at the start of surgery (T2), during mesh fixation (T3), and upon removal of the laryngeal mask (T4).

View Article and Find Full Text PDF

: Despite the known impact of propofol and remifentanil on hemodynamics and patient outcomes, there is a lack of comprehensive quantitative analysis, particularly in surgical settings, considering the influence of noxious stimuli. The aim of this study was to develop a quantitative semi-mechanistic population model that characterized the time course changes in mean arterial pressure (MAP) and heart rate (HR) due to the effects of propofol, remifentanil, and different types of noxious stimulation related to the clinical routine. : Data from a prospective study were used; the study analyzed the effects of propofol and remifentanil general anesthesia on female patients in physical status of I-II according to the American Society of Anesthesiologists (ASA I-II) undergoing gynecology surgery.

View Article and Find Full Text PDF

Want AI Summaries of new PubMed Abstracts delivered to your In-box?

Enter search terms and have AI summaries delivered each week - change queries or unsubscribe any time!