Kinetics of ethylene oxide desorption from sterilized materials.

J AOAC Int

Universidade Católica Portuguesa, CBQF/Centro de Biotecnologia e Química Fina, Escola Superior de Biotecnologia, R. Dr. António Bernardino de Almeida, 4200-072 Porto, Portugal.

Published: April 2013

Ethylene oxide gas is commonly used to sterilize medical devices, and concerns about using this agent on biological systems are well-established. Medical devices sterilized by ethylene oxide must be properly aerated to remove residual gas and by-products. In this work, kinetics of ethylene oxide desorption from different sterilized materials were studied in a range of aeration temperatures. The experimental data were well-described by a Fickian diffusion mass transfer behavior, and diffusivities were estimated for two textile and two polymeric materials within the temperature range of 1.5 to 59.0 degrees C. The results will allow predictions of ethylene oxide desorption, which is a key step for the design of sterilization/aeration processes, contributing to an efficient removal of residual ethylene oxide content.

Download full-text PDF

Source
http://dx.doi.org/10.5740/jaoacint.11-506DOI Listing

Publication Analysis

Top Keywords

ethylene oxide
24
oxide desorption
12
kinetics ethylene
8
desorption sterilized
8
sterilized materials
8
medical devices
8
oxide
6
ethylene
5
materials ethylene
4
oxide gas
4

Similar Publications

Want AI Summaries of new PubMed Abstracts delivered to your In-box?

Enter search terms and have AI summaries delivered each week - change queries or unsubscribe any time!