Purpose: Intramedullary nailing and locked plating for fixation of olecranon fractures has recently gained popularity. However, these two new technologies have not been compared for their biomechanical efficacy. The aim of this study was to evaluate the biomechanical stability of two newly designed fracture fixation devices for treating olecranon fractures during dynamic continuous loading: the ION intramedullary locking nail and the LCP precontoured locking compression plate.

Methods: Simulated oblique olecranon fractures were created in eight pairs of fresh-frozen cadaver ulnae and stabilised using either the LCP or ION. Specimens were then subjected to continuous dynamic loading (from 25 to 200 N), with a continuous angle alteration between 0° and 90° of flexion, to perform a matched-pairs comparison. Significant differences in the distance between markers surrounding the fracture gap was determined using the Wilcoxon test after four and 300 loading cycles.

Results: The ION resulted in significantly less displacement in the fracture gap at 0° extension (P = 0.036), 45° flexion (P = 0.035) and 90° flexion (P = 0.017) after 300 cycles of continuous loading. The measured displacements were small and were probably not of clinical significance. No mechanical failure or hardware migration was seen with either fixation technique.

Conclusion: This study shows significantly less micromotion for the ION than for the LCP in treating oblique olecranon fractures after 300 cycles of dynamic loading. Both implant types could be appropriate surgical techniques for fixation of selected olecranon fractures and osteotomies.

Download full-text PDF

Source
http://www.ncbi.nlm.nih.gov/pmc/articles/PMC3631502PMC
http://dx.doi.org/10.1007/s00264-013-1854-0DOI Listing

Publication Analysis

Top Keywords

olecranon fractures
24
intramedullary nailing
8
fixation olecranon
8
continuous loading
8
oblique olecranon
8
dynamic loading
8
90° flexion
8
fracture gap
8
300 cycles
8
olecranon
6

Similar Publications

Developing a Risk Score for Predicting Multiple Revision Surgeries in Patients With Fracture-Related Infections.

J Am Acad Orthop Surg

January 2025

From the Department of Orthopaedic Surgery, University of Alabama at Birmingham, Birmingham, AL (Yeager, Rutz, Strother, Spitler, and Johnson), and the Heersink School of Medicine, University of Alabama at Birmingham, Birmingham, AL (Gross, Benson, and Carter).

Introduction: Postoperative infections are a leading cause of morbidity following fracture repair. The purpose of this study is to develop a risk score predicting fracture-related infection (FRI) that will require one versus multiple revision surgeries related to infection eradication and bone healing.

Methods: This is a retrospective cohort study conducted at a single level I trauma center from 2013 to 2020.

View Article and Find Full Text PDF

Introduction: Olecranon fractures account for 5-7% of elbow fractures, making them a common injury. The majority of these fractures are treated surgically because they are intra-articular and can disrupt the elbow extensor mechanism. Mayo Type II fractures are displaced with a stable ulnohumeral joint, indicating intact ligamentous structures, particularly the anterior portion of the medial collateral ligament.

View Article and Find Full Text PDF

The traditional standard of care, tension band wire fixation modalities commonly used to treat non-comminuted olecranon fractures, are frequently associated with complications, with symptomatic hardware being the most common issue, often necessitating subsequent surgical procedures for hardware removal. We present a case of a young, active gentleman who sustained a simple olecranon fracture (Mayo type IIA) and underwent open reduction with the innovative all-suture non-metallic internal fixation procedure. We used a low-profile, cost-effective alternative fracture fixation technique, following the principles of pre-existing surgical techniques, while utilizing FiberWire sutures with biomechanical properties equivalent to, or possibly superior to, conventionally used metallic wires, thereby eliminating the need for additional surgical intervention.

View Article and Find Full Text PDF

Background: The financial and resource burden of management of olecranon fractures in the elderly is likely to increase with an aging population. There is limited evidence guiding treatment choice in this cohort. This study aimed to determine whether operative treatment of displaced olecranon fractures in elderly patients provides superior 12-month functional outcomes compared to nonoperative treatment.

View Article and Find Full Text PDF

Value driven outcome (VDO) assessment for the treatment of olecranon fractures.

Eur J Orthop Surg Traumatol

January 2025

Department of Orthopaedic Surgery, University of Utah, 590 Wakara Way, Salt Lake City, UT, 84108, USA.

Purpose: Controversy remains about the ideal construct for certain olecranon fractures. The purpose of this study was to compare cost-effectiveness with the value driven outcomes tool between fixation strategies of olecranon fractures.

Methods: All surgically treated isolated proximal ulna fractures (CPT code 24,685) at a level 1 trauma center from 2013 to 2023 were retrospectively reviewed.

View Article and Find Full Text PDF

Want AI Summaries of new PubMed Abstracts delivered to your In-box?

Enter search terms and have AI summaries delivered each week - change queries or unsubscribe any time!