Severity: Warning
Message: file_get_contents(https://...@pubfacts.com&api_key=b8daa3ad693db53b1410957c26c9a51b4908&a=1): Failed to open stream: HTTP request failed! HTTP/1.1 429 Too Many Requests
Filename: helpers/my_audit_helper.php
Line Number: 176
Backtrace:
File: /var/www/html/application/helpers/my_audit_helper.php
Line: 176
Function: file_get_contents
File: /var/www/html/application/helpers/my_audit_helper.php
Line: 250
Function: simplexml_load_file_from_url
File: /var/www/html/application/helpers/my_audit_helper.php
Line: 1034
Function: getPubMedXML
File: /var/www/html/application/helpers/my_audit_helper.php
Line: 3152
Function: GetPubMedArticleOutput_2016
File: /var/www/html/application/controllers/Detail.php
Line: 575
Function: pubMedSearch_Global
File: /var/www/html/application/controllers/Detail.php
Line: 489
Function: pubMedGetRelatedKeyword
File: /var/www/html/index.php
Line: 316
Function: require_once
Introduction: Maintaining contact with the patient during defibrillator discharge has been proposed as a method for reducing no flow time but carries an associated risk of electrocution of the rescuer. This study describes an investigation to determine if typical clinical examination gloves possess the dielectric strength needed to prevent breakdown at defibrillation voltages; a factor essential to protect the rescuer.
Methods: Four types of examination glove typically used in a clinical environment were tested with two types of defibrillation waveform commonly used. For each type of glove, 10 samples were tested initially using a monophasic defibrillation waveform and then, using a fresh sample of gloves, with a Biphasic waveform. For each glove the number of shocks required before electrical breakdown occurred was recorded.
Results: Kimberly Clark KC300 (nitrile), Kimberly Clark KC500 purple (nitrile), PH Medisavers GN90 (nitrile) and Bodyguards GL6622 (Vinyl) were tested using a monophasic defibrillation waveform and broke down after a median of 1, 4.5, 1 and 1 shocks respectively. The equivalent values for Biphasic defibrillator were 2, >10, 2.5 and 1 shocks.
Discussion: Typical clinical examination gloves do not possess the dielectric strength required to protect a rescuer from defibrillation voltages during hands-on chest compressions.