A brief history of LED photopolymerization.

Dent Mater

Faculty of Physics and Astronomy, Friedrich-Schiller-University Jena, Jena, Germany.

Published: June 2013

Objectives: The majority of modern resin-based oral restorative biomaterials are cured via photopolymerization processes. A variety of light sources are available for this light curing of dental materials, such as composites or fissure sealants. Quartz-tungsten-halogen (QTH) light curing units (LCUs) have dominated light curing of dental materials for decades and are now almost entirely replaced by modern light emitting diode light curing units (LED LCUs). Exactly 50 years ago, visible LEDs were invented. Nevertheless, it was not before the 1990s that LEDs were seriously considered by scientists or manufactures of commercial LCUs as light sources to photopolymerize dental composites and other dental materials. The objective of this review paper is to give an overview of the scientific development and state-of-the-art of LED photopolymerization of oral biomaterials.

Methods: The materials science of LED LCU devices and dental materials photopolymerized with LED LCU, as well as advantages and limits of LED photopolymerization of oral biomaterials, are discussed. This is mainly based on a review of the most frequently cited scientific papers in international peer reviewed journals. The developments of commercial LED LCUs as well as aspects of their clinical use are considered in this review.

Results: The development of LED LCUs has progressed in steps and was made possible by (i) the invention of visible light emitting diodes 50 years ago; (ii) the introduction of high brightness blue light emitting GaN LEDs in 1994; and (iii) the creation of the first blue LED LCUs for the photopolymerization of oral biomaterials. The proof of concept of LED LCUs had to be demonstrated by the satisfactory performance of resin based restorative dental materials photopolymerized by these devices, before LED photopolymerization was generally accepted. Hallmarks of LED LCUs include a unique light emission spectrum, high curing efficiency, long life, low energy consumption and compact device form factor.

Significance: By understanding the physical principles of LEDs, the development of LED LCUs, their strengths and limitations and the specific benefits of LED photopolymerization will be better appreciated.

Download full-text PDF

Source
http://dx.doi.org/10.1016/j.dental.2013.02.003DOI Listing

Publication Analysis

Top Keywords

led lcus
28
led photopolymerization
20
dental materials
20
light curing
16
led
13
light emitting
12
photopolymerization oral
12
light
10
lcus
9
light sources
8

Similar Publications

The Effects of Intensity, Exposure Time, and Distance of Polymerization Light on Vickers Microhardness and Temperature Rise of Conventional Resin-Based Composite.

Polymers (Basel)

November 2024

Department of Pediatric and Preventive Dentistry, Faculty of Dentistry with Dental Clinical Center, University of Sarajevo, 71000 Sarajevo, Bosnia and Herzegovina.

(1) Background: This study investigates the effects of curing light intensity, exposure time, and distance on the Vickers microhardness (VMH), hardness bottom-to-top ratio (HR), and temperature rise (TR) of conventional dental resin-based composite (RBC). (2) Materials and Methods: Specimens of one conventional RBC (Tetric EvoCeram, Ivoclar Vivadent) were cured with 12 different curing protocols (CPs), created with three different light intensities (Quartz Tungsten Halogen 300 mW/cm, LED 650 mW/cm, LED 1100 mW/cm), two exposure times (20 and 40 s), and two distances of curing tip (0 and 8 mm). The VMH of top (VMH-T) and bottom (VMH-B) surfaces was measured.

View Article and Find Full Text PDF

Objective: To evaluate the influence of battery level on power (mW), emission spectrum (mW/cm2/ nm), and light distribution on the active tip (mW/ cm2) of certified (FDA/ANVISA) and low-cost uncertified light-curing units (LCUs) purchased through e-commerce.

Methods: Seven LCUs, three certified: VALO Grand (Ultradent); Radii Xpert (SDI); and LED.B (Woodpecker); and four uncertified: 1 Sec; BS 300; LED curing light; and VAFU (VRN, AZDENT), were used.

View Article and Find Full Text PDF

Effect of duration and infection control barriers of light curing unit on hardness of Bulk Fill composite resin.

PeerJ

September 2024

Stomatology Hospital, School of Stomatology, Zhejiang University School of Medicine, Zhejiang Provincial Clinical Research Center for Oral Diseases, Key Laboratory of Oral Biomedical Research of Zhejiang Province, zhejiang university, Hangzhou, Zhejiang, China.

Background: This study aimed to investigate the impact of the duration of light curing unit (LCU) usage and the use of infection control barriers on the hardness of Bulk Fill composite resin after curing. The hypotheses were that extended usage of the LCU would not reduces its output power and resin hardness, and that the presence of polyethylene film barriers exacerbates the reduction in resin hardness.

Methods: Based on the absence or presence of polyethylene film (PE) and the number of layers used, a 3M LED curing light (EliparTM DeepCure-S; 3M ESPE, St Paul, MN, USA) was divided into three groups: PE0, PE1, and PE3.

View Article and Find Full Text PDF

The dark art of light curing in dentistry.

J Dent

November 2024

Department of Restorative Dentistry, State University of Ponta Grossa, Rua General Carlos Cavalcanti, 4748, Uvaranas, Ponta Grossa, 84030-900, Parana, Brazil.

Objective: This study was designed to show that the commonly reported irradiance values that are quoted in most publications inadequately describe the light output from light curing units (LCUs).

Methods: The total spectral radiant power (mW) output from 12 contemporary LCUs was measured with a fiberoptic spectroradiometer and a calibrated integrating sphere. Five recordings were taken for each LCU and exposure mode.

View Article and Find Full Text PDF

Background:  To evaluate the radiant power of the light cure units (LCUs) in relation to their type, radiant exitance, number of years in clinical use, and condition of LCUs tips in governmental and public clinics in Dental Faculties in Sana'a City.

Materials And Methods: LCUs were collected from different colleges at Sanaa City, Yemen, then LCU data as type, clinical age (<1 year, between 1-5 and ˃ 5-years), tip condition was visually inspected for damage and adhering debris, and the radiant exitance values of the tested LCUs. Radiant exitance values were subcategorized into three groups: <400, 400-850, and >850 mW/cm², labeled as inadequate, marginal, and adequate radiant exitances, respectively.

View Article and Find Full Text PDF

Want AI Summaries of new PubMed Abstracts delivered to your In-box?

Enter search terms and have AI summaries delivered each week - change queries or unsubscribe any time!