This study compared the Biomic automated well reader results to the MicroScan WalkAway results for reading MicroScan antimicrobial susceptibility and identification panels at four different sites. Routine fresh clinical isolates and quality control (QC) organisms were tested at each study site. A total of 46,176 MicroScan panel drug-organism combinations were read. The Biomic category agreement for 3,117 Gram-negative bacteria was 98.4%, with 1.4% minor and 0.2% major discrepancies. The Biomic category agreement for 5,233 Gram-positive bacteria was 98.7%, with 0.9% minor, 0.3% major, and 0.1% very major errors. Essential agreement, defined as Biomic results that were within ±1 2-fold dilution of the MicroScan results, was 99.3% for Gram-negative bacteria and 98.3% for Gram-positive bacteria. Biomic reading of MicroScan identification panels provided an overall agreement (first- and second-choice organism match) of 99.5% with 846 Gram-negative isolates and 99.5% with 430 Gram-positive isolates. These results suggest that the Biomic automated reader can provide accurate reading of MicroScan panels and has the capability of a visual panel read for manual adjustment of results.

Download full-text PDF

Source
http://www.ncbi.nlm.nih.gov/pmc/articles/PMC3647896PMC
http://dx.doi.org/10.1128/JCM.03088-12DOI Listing

Publication Analysis

Top Keywords

reading microscan
16
biomic automated
12
identification panels
12
microscan
8
microscan walkaway
8
walkaway reading
8
microscan antimicrobial
8
antimicrobial susceptibility
8
susceptibility identification
8
biomic category
8

Similar Publications

Want AI Summaries of new PubMed Abstracts delivered to your In-box?

Enter search terms and have AI summaries delivered each week - change queries or unsubscribe any time!