Accuracy of carboxyhemoglobin detection by pulse CO-oximetry during hypoxemia.

Anesth Analg

From the Department of Anesthesia and Perioperative Care, University of California, San Francisco, California.

Published: October 2013

Background: Carbon monoxide poisoning is a significant problem in most countries, and a reliable method of quick diagnosis would greatly improve patient care. Until the recent introduction of a multiwavelength "pulse CO-oximeter" (Masimo Rainbow SET(®) Radical-7), obtaining carboxyhemoglobin (COHb) levels in blood required blood sampling and laboratory analysis. In this study, we sought to determine whether hypoxemia, which can accompany carbon monoxide poisoning, interferes with the accurate detection of COHb.

Methods: Twelve healthy, nonsmoking, adult volunteers were fitted with 2 standard pulse-oximeter finger probes and 2 Rainbow probes for COHb detection. A radial arterial catheter was placed for blood sampling during 3 interventions: (1) increasing hypoxemia in incremental steps with arterial oxygen saturations (SaO2) of 100% to 80%; (2) normoxia with incremental increases in %COHb to 12%; and (3) elevated COHb combined with hypoxemia with SaO2 of 100% to 80%. Pulse-oximeter (SpCO) readings were compared with simultaneous arterial blood values at the various increments of hypoxemia and carboxyhemoglobinemia (≈25 samples per subject). Pulse CO-oximeter performance was analyzed by calculating the mean bias (SpCO - %COHb), standard deviation of the bias (precision), and the root-mean-square error (A(rms)).

Results: The Radical-7 accurately detected hypoxemia with both normal and elevated levels of COHb (bias mean ± SD: 0.44% ± 1.69% at %COHb <4%, and -0.29% ± 1.64% at %COHb ≥4%, P < 0.0001, and A(rms) 1.74% vs 1.67%). COHb was accurately detected during normoxia and moderate hypoxia (bias mean ± SD: -0.98 ± 2.6 at SaO2 ≥95%, and -0.7 ± 4.0 at SaO2 <95%, P = 0.60, and A(rms) 2.8% vs 4.0%), but when SaO2 decreased below approximately 85%, the pulse CO-oximeter always gave low signal quality errors and did not report SpCO values.

Conclusions: In healthy volunteers, the Radical-7 pulse CO-oximeter accurately detects hypoxemia with both low and elevated COHb levels, and accurately detects COHb, but only reads SpCO when SaO2 is more than approximately 85%.

Download full-text PDF

Source
http://www.ncbi.nlm.nih.gov/pmc/articles/PMC4476500PMC
http://dx.doi.org/10.1213/ANE.0b013e31828610a0DOI Listing

Publication Analysis

Top Keywords

carbon monoxide
8
monoxide poisoning
8
blood sampling
8
sao2 100%
8
100% 80%
8
hypoxemia
6
accuracy carboxyhemoglobin
4
carboxyhemoglobin detection
4
detection pulse
4
pulse co-oximetry
4

Similar Publications

Background: Data regarding the effectiveness and safety of endoscopic lung volume reduction with valves (ELVR) in emphysema patients with a very low 6-min walk test (6MWT) are limited. Patients with severe emphysema and very low exercise capacity, as indicated by a 6MWT ≤140 m, are often excluded from clinical studies on ELVR, assuming limited therapeutic benefits and increased complication risk.

Study Designs And Methods: This study utilised data from the Lungenemphysemregister e.

View Article and Find Full Text PDF

Association between impaired diffusion capacity and small airway dysfunction: a cross-sectional study.

ERJ Open Res

January 2025

State Key Laboratory of Respiratory Disease & National Clinical Research Center for Respiratory Disease & Guangzhou Institute of Respiratory Health & National Center for Respiratory Medicine & Department of Pulmonary and Critical Care Medicine, The First Affiliated Hospital of Guangzhou Medical University, Guangzhou Medical University, Guangzhou, Guangdong, China.

Background: Small airway dysfunction (SAD) and impaired diffusion capacity of the lungs for carbon monoxide ( ) are positively associated with a worse prognosis. Individuals with both dysfunctions have been identified in clinical practice and it is unknown whether they have worse health status or need management. We conducted this study to explore the association between SAD and impaired , and the difference between the groups with two dysfunctions, with either one dysfunction and with no dysfunction.

View Article and Find Full Text PDF

This work describes the design and implementation of optics for EXCLAIM, the EXperiment for Cryogenic Large-Aperture Intensity Mapping. EXCLAIM is a balloon-borne telescope that will measure integrated line emission from carbon monoxide at redshifts z < 1 and ionized carbon ([CII]) at redshifts z = 2.5 - 3.

View Article and Find Full Text PDF

Repeated measurements of household air pollution may provide better estimates of average exposure but can add to costs and participant burden. In a randomized trial of gas versus biomass cookstoves in four countries, we took supplemental personal 24-h measurements on a 10% subsample for mothers and infants, interspersed between protocol samples. Mothers had up to five postrandomization protocol measurements over 16 months, while infants had three measurements over one year.

View Article and Find Full Text PDF

Introduction: Delayed neurological sequelae is a common complication following carbon monoxide poisoning, which significantly affects the quality of life of patients with the condition. We aimed to develop a machine learning-based prediction model to predict the frequency of delayed neurological sequelae in patients with carbon monoxide poisoning.

Methods: A single-center retrospective analysis was conducted in an emergency department from January 01, 2018, to December 31, 2023.

View Article and Find Full Text PDF

Want AI Summaries of new PubMed Abstracts delivered to your In-box?

Enter search terms and have AI summaries delivered each week - change queries or unsubscribe any time!