Background: Fractures of pace/sense conductors in implantable cardioverter-defibrillator (ICD) leads have been studied extensively, but little is known about fractures of high-voltage (HV) conductors.

Objective: To characterize the presentation of isolated HV conductor fractures, define the optimal impedance threshold for identifying them, and compare it to the existing nominal impedance threshold (200 Ω) for patient and remote-monitoring alerts.

Methods: This retrospective study analyzed HV fractures in explanted, dual-coil, model 6949 Sprint Fidelis leads (Medtronic, Minneapolis, MN). The study group consisted of 25 leads with structurally and electrically confirmed HV conductor fractures; 41 leads that were structurally and electrically intact served as controls. We analyzed long-term HV impedance trends from stored ICD data files of both groups to determine the optimal impedance threshold that would discriminate fractures from normal leads.

Results: In the study group, 14 leads (56%) had fractures of the cable to the right ventricular coil, 9 (36 %) leads had fractures of the cable to the superior vena cava (SVC) coil, and 2 (8%) had both. We found that an impedance threshold of >100 Ω and/or an abrupt 75% increase in chronic HV impedance were diagnostic of HV conductor fractures with 100% sensitivity and specificity. HV fractures proximal to the SVC coil were more likely to be associated with concomitant pace/sense fractures. Large (200 Ω to infinity), abrupt increases in impedance were more common when fractures occurred proximal to the right ventricular coil but distal to the SVC coil.

Conclusions: HV conductor fractures can be diagnosed when HV impedance exceeds 100 Ω or abruptly increases by 75% from baseline.

Download full-text PDF

Source
http://dx.doi.org/10.1016/j.hrthm.2013.02.027DOI Listing

Publication Analysis

Top Keywords

conductor fractures
20
impedance threshold
16
fractures
14
sprint fidelis
8
fidelis leads
8
impedance
8
optimal impedance
8
study group
8
leads structurally
8
structurally electrically
8

Similar Publications

In-Situ Cross-Linked Polymers for Enhanced Thermal Cycling Stability in Flexible Perovskite Solar Cells.

Angew Chem Int Ed Engl

December 2024

Key Lab for Special Functional Materials of Ministry of Education, National & Local Joint Engineering Research Center for High-efficiency Display and Lighting Technology, School of Nanoscience and Materials Engineering, Collaborative Innovation Center of Nano Functional Materials and Applications, Henan University, Kaifeng, 475004, P. R. China.

Flexible perovskite solar cells (FPSCs) are a promising emerging photovoltaic technology, with certified power conversion efficiencies reaching 24.9 %. However, the frequent occurrence of grain fractures and interface delamination raises concerns about their ability to endure the mechanical stresses caused by temperature fluctuations.

View Article and Find Full Text PDF

Stretchable supercapacitors are essential components in wearable electronics due to their low heat generation and seamless integration capabilities. Thermoplastic polyurethane elastomers, recognized for their dynamic hydrogen-bonding structure, exhibit excellent stretchability, making them well-suited for these applications. This study introduces fluorine-based interactions in the hard segments of thermoplastic polyurethanes, resulting in polyurethanes with a low elastic modulus, high fracture strength, exceptional fatigue resistance, and self-healing properties.

View Article and Find Full Text PDF

Gel with ionic conductivity and stretchability is considered as an ideal alternative to conventional rigid metallic conductors in the flexible electronics. However, present gels suffer from poor mechanical properties and crack sensitivity due to their weak intermolecular (chain) interactions and homogeneous network structure. Herein, a transparent and tough polyacrylamide (PAM) ionogel is designed, which can form stress-induced microphase-separated domains with high hydrogen bonding density under stress to inhibit crack propagation.

View Article and Find Full Text PDF

Background: Left bundle branch area pacing (LBBAP) has gained increased adoption globally with the use of both lumenless leads (LLLs) and stylet-driven leads (SDLs). As these leads have been developed for conventional endocardial pacing sites, concerns remain regarding the lead integrity with LBBAP.

Objectives: This study evaluates lead integrity of pacing leads used for LBBAP in a large, real-world cohort of patients with LBBAP.

View Article and Find Full Text PDF

Human skin is essential for perception, encompassing haptic, thermal, proprioceptive, and pain-sensing functions through ion movement. Additionally, it is mechanically resilient and self-healing for protection. Inspired by these unique properties, researchers have attempted to develop stretchable, self-healing sensors based on ion dynamics.

View Article and Find Full Text PDF

Want AI Summaries of new PubMed Abstracts delivered to your In-box?

Enter search terms and have AI summaries delivered each week - change queries or unsubscribe any time!