Background: In patients with patent ductus arteriosus (PDA), transcatheter closure is the current procedure of choice. There are multiple devices available with limited current comparative data and varied recommendations for device selection.

Objective: To assess the efficacy and safety of the Flipper coil (FC) and the Amplatzer Duct Occluder (ADO).

Methods: An intention to treat analysis of all children admitted to the catheter laboratory at a single institution for occlusion of PDA from 2003 to 2011 was performed. Patient and device selection were determined by the treating physician. Standard techniques for FC and ADO implantation were used.

Results: Two hundred and twenty eight children with median weight of 14.2 kg (range; 5.5-68 kg) underwent cardiac catheterisation, with successful occlusion in 96.2% of patients when attempted. In 16 patients, median angiographic PDA diameter of 0.8 mm (range; 0.4-1.2 mm), was considered too small to warrant closure. Eight patients with large PDA's underwent surgical ligation. FC was successfully used in 70 (34.3%) and ADO in 134 (66.7%) patients. ADO patients were smaller (P=0.004) with larger PDA's (P<0.0001) than the FC group. Median fluoroscopy time was longer for ADO patients (10.1 min vs 8.0 min; P<0.0001). ADO had a lower embolisation rate (0% vs 6.6%; P=0.005) and a higher complete occlusion rate at follow-up (100% vs 73.4%; P<0.0001). Length of hospital admission decreased with time in both groups.

Conclusion: Transcatheter closure of the PDA has a high degree of safety and efficacy. This study suggests that the ADO may be the device of first choice in the current era.

Download full-text PDF

Source
http://dx.doi.org/10.1016/j.hlc.2012.12.007DOI Listing

Publication Analysis

Top Keywords

transcatheter closure
8
patent ductus
8
ductus arteriosus
8
intention treat
8
treat analysis
8
patients
6
closure patent
4
arteriosus intention
4
analysis background
4
background patients
4

Similar Publications

Background: Transcatheter closure of large and complex atrial septal defect can pose challenges and complications during device placement. To improve stability, several assistive techniques have been developed.

Methods: This retrospective study evaluated the efficacy of the device-assisted device closure technique for large secundum atrial septal defects.

View Article and Find Full Text PDF

Transcatheter closure (TCC) of certain ventricular septal defect (VSD) subtypes typically requires arteriovenous loop (AVL) formation or retrograde transarterial deployment. Upfront transvenous cannulation from the right ventricle avoids arterial access and loop-related complications. We retrospectively reviewed data of patients who underwent TCC for perimembranous, intraconal, and post-surgical residual VSDs at our institution (January 2019-December 2023).

View Article and Find Full Text PDF

Want AI Summaries of new PubMed Abstracts delivered to your In-box?

Enter search terms and have AI summaries delivered each week - change queries or unsubscribe any time!