Purpose: A novel highly accurate method for MR thermometry, effective at high field, is introduced and validated, which corrects for slow and fast field fluctuations by means of reference images.

Methods: An asymmetric spin-echo echo planar imaging sequence was made frequency-selective to water or a reference substance by controlling the slice-select gradient polarity and the duration of the excitation and refocusing radiofrequency pulses. Images were acquired pairwise, and the temperature-sensitive water images were corrected for field fluctuations using the reference images. In a phantom radiofrequency heating experiment, dissolved dimethyl sulfoxide was used as a reference substance. Temperature stability was tested in vivo on the human brain, referenced using subcutaneous scalp fat. Water and fat phase images were acquired only 50 ms apart. Bloch simulations validated the frequency selection accuracy.

Results: Asymmetric spin-echo imaging using a simple frequency selection method provides highly accurate referenced MR thermometry in phantoms and in vivo at 7 T. Effects of field fluctuations caused by field drift, breathing, and heart beat were corrected. The technique is highly robust against B1 inhomogeneities.

Conclusion: Frequency selection using gradient-reversal can enable fast accurate referenced in vivo MR thermometry, assisting thermal characterization of radiofrequency coils and possibly in vivo SAR monitoring.

Download full-text PDF

Source
http://dx.doi.org/10.1002/mrm.24681DOI Listing

Publication Analysis

Top Keywords

asymmetric spin-echo
12
field fluctuations
12
frequency selection
12
fast accurate
8
high field
8
highly accurate
8
fluctuations reference
8
reference substance
8
images acquired
8
accurate referenced
8

Similar Publications

Want AI Summaries of new PubMed Abstracts delivered to your In-box?

Enter search terms and have AI summaries delivered each week - change queries or unsubscribe any time!