Aims: For most elderly pacemaker patients, evaluation of rate-adaptive pacing using treadmill and bicycle tests is impractical and not representative of typical daily activities. This study was designed to compare the performance and physiological response of the closed-loop stimulation (CLS) rate-adaptive sensor to accelerometer (XL) and no rate sensor (DDD) during typical daily activity testing.
Methods And Results: Subjects recently implanted with a Cylos pacemaker completed timed activities of daily life testing, which included walking, sweeping, and standing from a seated position. Activity performance and physiological response from each sensor mode was evaluated for subjects requiring ≥80% pacing. Overall, 74 subjects needed ≥80% pacing during at least one test. An increase in the area swept (CLS vs. XL, 1.67 m(2) difference, P = 0.009; CLS vs. DDD, 1.59 m(2) difference, P = 0.025) and a decrease in the prevalence of orthostatic hypotension (OH) after standing 1 min (CLS vs. XL, odds ratio = 0.16, P = 0.006; CLS vs. DDD, odds ratio = 0.18, P = 0.012) was observed in the CLS mode as compared with XL and DDD. No statistical difference in walk distance was observed between CLS and XL or CLS and DDD.
Conclusion: In acute testing, as compared with XL and DDD, CLS provides a more physiological response during the performance of activities of daily living for subjects with ≥80% pacing. This is clinically reflected in better performance during the sweep test as well as a decrease in the prevalence of OH in our elderly population. Clinicaltrials.gov identifier: NCT00355797.
Download full-text PDF |
Source |
---|---|
http://www.ncbi.nlm.nih.gov/pmc/articles/PMC3663333 | PMC |
http://dx.doi.org/10.1093/europace/eus425 | DOI Listing |
Enter search terms and have AI summaries delivered each week - change queries or unsubscribe any time!