The correlation between retinal blood flow velocity measured by the retinal function imager and various physiological parameters.

Ophthalmic Surg Lasers Imaging Retina

Department of Ophthalmology, The Edith Wolfson Medical Center, Sackler Faculty of Medicine, Tel Aviv University, Holon, Israel.

Published: July 2013

Background And Objective: The retinal function imager (RFI) (Optical Imaging Ltd., Rehovot, Israel) measures retinal blood flow velocity non-invasively. The authors studied the reproducibility of these measurements and assessed the effect of physiological components on them.

Patients And Methods: Sixty-seven individuals with no retinal pathology were recruited. Velocity reproducibility was verified by comparing repeated RFI measurements. The correlation of the velocity with physiological parameters was assessed by mixed linear and Gaussian models.

Results: The average velocity was 4.2 ± 0.9 mm/sec arterial and 3.3 ± 0.8 mm/sec venous. Variability was 7.5% ± 3.7% and interclass correlation coefficient was r = 0.744. Venous velocity decreased after 40 years of age (0.32 mm/sec per decade, P < .01). Arterial velocity increased as mean arterial pressure increased (0.25 mm/sec per 10 mm Hg, P < .01). There was also a positive association between velocities and heart rate (arteries: 0.21 mm/sec per 10 bpm, P < .05; veins: 0.22 mm/sec per 10 bpm, P < .01).

Conclusion: The RFI provides a reproducible, non-invasive technique to assess retinal velocities.

Download full-text PDF

Source
http://dx.doi.org/10.3928/23258160-20121221-13DOI Listing

Publication Analysis

Top Keywords

retinal blood
8
blood flow
8
flow velocity
8
retinal function
8
function imager
8
physiological parameters
8
mm/sec bpm
8
velocity
7
mm/sec
6
retinal
5

Similar Publications

Want AI Summaries of new PubMed Abstracts delivered to your In-box?

Enter search terms and have AI summaries delivered each week - change queries or unsubscribe any time!