A PHP Error was encountered

Severity: Warning

Message: file_get_contents(https://...@pubfacts.com&api_key=b8daa3ad693db53b1410957c26c9a51b4908&a=1): Failed to open stream: HTTP request failed! HTTP/1.1 429 Too Many Requests

Filename: helpers/my_audit_helper.php

Line Number: 176

Backtrace:

File: /var/www/html/application/helpers/my_audit_helper.php
Line: 176
Function: file_get_contents

File: /var/www/html/application/helpers/my_audit_helper.php
Line: 250
Function: simplexml_load_file_from_url

File: /var/www/html/application/helpers/my_audit_helper.php
Line: 3122
Function: getPubMedXML

File: /var/www/html/application/controllers/Detail.php
Line: 575
Function: pubMedSearch_Global

File: /var/www/html/application/controllers/Detail.php
Line: 489
Function: pubMedGetRelatedKeyword

File: /var/www/html/index.php
Line: 316
Function: require_once

Aortic valve hydrodynamics: considerations on the absence of sinuses of Valsalva. | LitMetric

AI Article Synopsis

  • The study aimed to compare the hydrodynamics of the CEM and SJME heart valves at varying stroke volumes and pulse rates in two different aortic conduits.
  • The testing involved measuring systolic and diastolic performances, leaflet coaptation time, and maximum flow rates at different heart rates and stroke volumes.
  • Results showed that CEM valves had lower gradients and larger effective orifice areas compared to SJME valves, and the absence of sinuses of Valsalva affected the valve performance, potentially increasing stress on the valve tissue.

Article Abstract

Background And Aim Of The Study: The study aim was to compare the hydrodynamics of the Carpentier-Edwards Magna 21 (CEM) and St. Jude Medical Biocor-Epic-Supra 21 (SJME) valves at increasing stroke volume and pulse rate in two different aortic conduits, namely straight and with sinuses of Valsalva present.

Methods: Both valve types were tested in the aortic chamber of the Sheffield pulse duplicator, at rates of 70, 80 and 90 beats/min, and stroke volumes of 50 and 60 ml. The systolic and diastolic performances were each recorded. The leaflet coaptation time, ventricle isovolumetric time and maximum instantaneous flow rate were also recorded.

Results: Regardless of the aortic conduit, CEM valves showed a significantly lower gradient than SJME valves (p < 0.05), and a significantly larger effective orifice area (EOA) (p < 0.05); the latter parameter was unaffected for both valves, at an increasing pulse rate (p > 0.05). The maximum transvalvular flow velocity was significantly higher in the straight conduit for both valves (p < 0.05). With regards to diastole, the SJME valve showed the lowest regurgitant volume (p < 0.05). The leaflet coaptation time was significantly shorter for the SJME valve than for the CEM valve (p < 0.05), but when tested in a straight conduit it was shortened significantly for both valves (p < 0.05).

Conclusion: An absence of the sinuses of Valsalva may modify the diastolic and systolic behaviors of the tissue valve leaflets by reducing the time required for leaflet coaptation, and increasing the valve closing volume and maximum transvalvular flow velocity. It is speculated that these hydrodynamic changes may increase the working stress on the valve tissue, leading to possible premature structural valve deterioration.

Download full-text PDF

Source

Publication Analysis

Top Keywords

sinuses valsalva
12
leaflet coaptation
12
absence sinuses
8
sjme valves
8
valves increasing
8
pulse rate
8
valve
8
coaptation time
8
valves 005
8
maximum transvalvular
8

Similar Publications

Want AI Summaries of new PubMed Abstracts delivered to your In-box?

Enter search terms and have AI summaries delivered each week - change queries or unsubscribe any time!