A 22-year old man presented with a massive haemothorax 25 days after bullectomy for a spontaneous pneumothorax. Thoracoscopic surgery revealed ongoing bleeding from the chest wall caused by a sharp edge of the Endoloop Ligature (Ethicon Endo-Surgery) used to resect the remaining small part of the lung at the earlier staple bullectomy. The point where bleeding was occurring was clipped and covered using a collagen patch coated with human fibrinogen and thrombin. The protruding sharp edge of the Endoloop was excised together with the surrounding lung tissue, using a stapler. Although prevention of this type of complication is difficult, awareness of the potential problem may help in managing such extremely rare events.
Download full-text PDF |
Source |
---|---|
http://www.ncbi.nlm.nih.gov/pmc/articles/PMC3630437 | PMC |
http://dx.doi.org/10.1093/icvts/ivt047 | DOI Listing |