Introduction: This study aimed to investigate in an ex vivo model the reduction in patient radiation dose while maintaining accurate linear measurements by comparing cone-beam computed tomography (CBCT) scans taken at 360° versus 180° rotation, with porcine jaw specimens as a reference standard.

Methods: CBCT scans of 12 sectioned porcine hemimandibles at 360° and 180° rotations were taken with standardized clinical exposure parameters. To assess interobserver variability, 6 assessors who were blinded to the degree of rotation took linear measurements of anatomic structures on each scan. The measurements were repeated after 2 weeks to assess intraobserver variability. Accuracy of measurement was judged against the corresponding measurements taken from the porcine jaw specimens.

Results: Intraclass correlations signaled good-to-excellent intraobserver and interobserver agreement (0.65-0.98 and 0.79-0.98), respectively. Mixed regression analysis found no significant difference between the measurements from 180° or 360° rotations and no difference between the 2 rotations and porcine jaw specimens.

Conclusions: A CBCT image sufficient to make accurate clinical measurements with a reduced radiation exposure may be obtained by using 180° rotation of the CBCT tube head.

Download full-text PDF

Source
http://dx.doi.org/10.1016/j.joen.2012.11.017DOI Listing

Publication Analysis

Top Keywords

porcine jaw
12
cone-beam computed
8
computed tomography
8
linear measurements
8
cbct scans
8
180° rotation
8
measurements
7
vitro comparison
4
comparison accuracy
4
accuracy measurements
4

Similar Publications

Want AI Summaries of new PubMed Abstracts delivered to your In-box?

Enter search terms and have AI summaries delivered each week - change queries or unsubscribe any time!