Purpose: Phosphorous metabolite ratios in human myocardium were determined by a combination of acquisition weighted CSI and a SLOOP evaluation and the results were compared to corresponding SLOOP experiments using standard CSI.
Materials And Methods: 10 healthy subjects were examined at 1.5 T using both standard CSI and acquisition weighted CSI. Both experiments were performed with a similar acquisition time and the same spatial resolution. The PCr/ATP ratio was determined and the localization properties of both experiments were compared.
Results: The PCr/ATP ratio of 2.2±0.4 found for the experiment using acquisition weighted CSI was almost identical to the value of 2.0±0.4 for standard CSI. The sensitivity and the localization properties improved in all subjects using SLOOP evaluation of the acquisition weighted sampling in comparison to the standard CSI acquisition with an average of 3% and 18%, respectively.
Conclusion: The employment of acquisition weighting allows for a further improvement of the (31)P SLOOP spectroscopy of the human heart.
Download full-text PDF |
Source |
---|---|
http://dx.doi.org/10.1016/j.zemedi.2013.01.002 | DOI Listing |
In the realm of 3D measurement, photometric stereo excels in capturing high-frequency details but suffers from accumulated errors that lead to low-frequency distortions in the reconstructed surface. Conversely, light field (LF) reconstruction provides satisfactory low-frequency geometry but sacrifices spatial resolution, impacting high-frequency detail quality. To tackle these challenges, we propose a photometric stereoscopic light field measurement (PSLFM) scheme that harnesses the strengths of both methods.
View Article and Find Full Text PDFSci Rep
January 2025
Centre for Applied Photonics, INESC TEC, Rua do Campo Alegre 687, 4169-007, Porto, Portugal.
Spectral Imaging techniques such as Laser-induced Breakdown Spectroscopy (LIBS) and Raman Spectroscopy (RS) enable the localized acquisition of spectral data, providing insights into the presence, quantity, and spatial distribution of chemical elements or molecules within a sample. This significantly expands the accessible information compared to conventional imaging approaches such as machine vision. However, despite its potential, spectral imaging also faces specific challenges depending on the limitations of the spectroscopy technique used, such as signal saturation, matrix interferences, fluorescence, or background emission.
View Article and Find Full Text PDFMagn Reson Med
January 2025
Department 8.1 - Biomedical Magnetic Resonance, Physikalisch-Technische Bundesanstalt (PTB), Braunschweig and Berlin, Germany.
Purpose: To develop a low-cost, high-performance, versatile, open-source console for low-field MRI applications that can integrate a multitude of different auxiliary sensors.
Methods: A new MR console was realized with four transmission and eight reception channels. The interface cards for signal transmission and reception are installed in PCI Express slots, allowing console integration in a commercial PC rack.
Magn Reson Imaging
January 2025
Department of Radiology, Albert Einstein College of Medicine and Montefiore Medical Center, Bronx, NY, USA. Electronic address:
Purpose: Diffusion-weighted arterial spin labeling (DW-ASL) MRI has been proposed to determine the rate of water exchange (K) across the blood brain barrier (BBB). This study aims to further evaluate K MRI by comparing it with standard dynamic contrast-enhanced (DCE) MRI and histology in association with mannitol-induced disruption of the BBB.
Methods: DW-ASL was measured using a multiple b-value MRI protocol in normal rats at three post-labeling delays (N = 19), before and after intra-carotid injection of mannitol to disrupt BBB in one hemisphere (N = 13).
Sensors (Basel)
January 2025
College of Science and Technology, Ningbo University, Ningbo 315000, China.
The quality of surface morphology can reflect the electrical performance of silver-based contacts. Existing research on the correlation of morphological-electrical performance is based solely on empirical models from traditional visual inspections and only considers the impact of visually observable macro-textural features on electrical performance. However, the influence of micro-textural features on electrical performance should not be overlooked.
View Article and Find Full Text PDFEnter search terms and have AI summaries delivered each week - change queries or unsubscribe any time!