Purpose: Many studies have demonstrated higher precision and better radiological results in Total knee arthroplasty (TKA) with computer-assisted surgery (CAS). On the other hand, studies revealed a lengthening of operation time up to 20 min for this technique and demonstrated rare additional complications as fractures and neurovascular injuries caused by the array pins and any intraoperative array dislocation leads to abortion of CAS. To combine the advantages and eliminate the disadvantages of standard CAS, we evaluated the accuracy of a so-called pinless CT-free version of knee navigation (pinless CAS) abandoning the reference pins and reducing the necessary workflow to a minimum.
Method: The present study compares the accuracy of the reference methods of two different CT-free knee navigation software versions (Brainlab Knee 2.1 and Brainlab Knee Express 2.5). Thirty patients received TKA assisted by standard CAS. Intraoperatively, the proposed bony resections of standard CAS were matched with the new pinless CAS. Postoperatively, the results were checked by evaluating the radiographs concerning leg axis, femoral flexion and tibial slope.
Results: All results concerning precise cuts (femoral as well as tibial coronal/varus-valgus alignment, femoral flexion alignment and tibial slope, resection height) were comparable between both groups (n.s.). In femoral, we found a mean deviation of coronal alignment of 0.3° (SD 0.7) and flexion of 0.2° (SD 0.8). In tibial, we found a mean deviation of coronal alignment of 0.2° (SD 0.5) and slope of 0.2° (SD 0.6). The mean additional operation time for the pinless CAS was below 2 min. The postoperative mechanical leg axis was within the threshold of 3° in all patients, tibial slope and femoral flexion matched with CAS values.
Conclusion: In clinical routine, pinless CAS can comprise the advantages of CAS leaving the disadvantages aside. It reduces surgical time and avoids complications associated with the tracking pins of conventional CAS.
Download full-text PDF |
Source |
---|---|
http://dx.doi.org/10.1007/s00167-013-2430-2 | DOI Listing |
J Robot Surg
August 2022
Zimmer CAS, Montréal, QC, H3C-2N6, Canada.
Accurate component orientation and restoration of hip biomechanics remains a continuing challenge in total hip arthroplasty (THA). The goal of this study was to analyze the accuracy/reproducibility of a novel CT-free and pin-less robotic-assisted THA (RA-THA) platform compared to manual THA (mTHA). This matched-pair cadaveric study compared this RA-THA system to mTHA (n = 33/arm), both using the assistance of fluoroscopic imaging, in a group of 14 high-volume arthroplasty surgeons.
View Article and Find Full Text PDFJ Orthop Surg (Hong Kong)
January 2017
Department of Orthopaedic Surgery, Singapore General Hospital, Singapore, Singapore.
This study aims to evaluate the 2-year post-operative clinical outcomes of patients undergoing total knee arthroplasty (TKA) with computer-assisted surgery (CAS) using the pinless BrainLAB VectorVision Knee 2.5 Navigation System versus standard CAS. A retrospective case-control study of 200 patients who underwent TKA with CAS from 2008 to 2012 was conducted.
View Article and Find Full Text PDFEur J Orthop Surg Traumatol
January 2017
The South West London Elective Orthopaedic Centre, Dorking Road, Epsom, London, KT18 7EG, UK.
Purpose: Accurate component alignment in total knee replacement (TKR) is one of the important factors in determining long-term survivorship. This has been achieved by conventional jigs, computer-assisted technology (CAS) and more recently patient-specific instrumentation (PSI). The purpose of the current study was to investigate the in vivo accuracy of Trumatch™ PSI using validated pin-less computer navigation system.
View Article and Find Full Text PDFKnee Surg Sports Traumatol Arthrosc
August 2014
Orthopaedic Department, University of Regensburg, Kaiser-Karl V.-Allee 3, 93077, Bad Abbach, Germany,
Purpose: Many studies have demonstrated higher precision and better radiological results in Total knee arthroplasty (TKA) with computer-assisted surgery (CAS). On the other hand, studies revealed a lengthening of operation time up to 20 min for this technique and demonstrated rare additional complications as fractures and neurovascular injuries caused by the array pins and any intraoperative array dislocation leads to abortion of CAS. To combine the advantages and eliminate the disadvantages of standard CAS, we evaluated the accuracy of a so-called pinless CT-free version of knee navigation (pinless CAS) abandoning the reference pins and reducing the necessary workflow to a minimum.