Accelerated MRI by SPEED with generalized sampling schemes.

Magn Reson Med

Institute of Information and Control, Hangzhou Dianzi University, Hangzhou, Zhejiang, People's Republic of China.

Published: December 2013

Purpose: To enhance the fast imaging technique of skipped phase encoding (PE) and edge deghosting (SPEED) for more general sampling options, and thus more flexibility in implementations and applications.

Methods: SPEED uses skipped PE steps to accelerate MRI scan. Previously, the PE skip size was chosen from prime numbers only. This restriction has been relaxed in this study to allow choice of any integers rather than merely prime numbers. Various sampling patterns were studied under all possible combinations of PE skip size and PE shifts. A criterion based on the rank values of ghost phasor matrices was introduced to evaluate SPEED reconstruction.

Results: The reconstruction quality was found to correlate with the rank value of the ghost phasor matrix and the skipped PE size N. A low-rank value indicates a singular matrix that causes failure of the SPEED reconstruction. Composite numbers combined with appropriately chosen PE shifts yielded satisfactory reconstruction results.

Conclusion: With properly chosen PE shifts, it was found that any integers, including both prime numbers and composite numbers, could be used as PE skip size for SPEED. This finding allows much more flexible data acquisition options that may lead to more freedom in practical implementations and applications.

Download full-text PDF

Source
http://dx.doi.org/10.1002/mrm.24605DOI Listing

Publication Analysis

Top Keywords

skip size
12
prime numbers
12
ghost phasor
8
composite numbers
8
chosen shifts
8
speed
6
numbers
5
accelerated mri
4
mri speed
4
speed generalized
4

Similar Publications

Want AI Summaries of new PubMed Abstracts delivered to your In-box?

Enter search terms and have AI summaries delivered each week - change queries or unsubscribe any time!