Background And Aim: A current drawback of endoscopic submucosal dissection (ESD) for early-stage gastrointestinal tumors is the lack of instruments that can safely assist with this procedure. We have developed a pulsed jet device that can be incorporated into a gastrointestinal endoscope. Here, we investigated the mechanical profile of the pulsed jet device and demonstrated the usefulness of this instrument in esophageal ESD in swine.

Methods: The device comprises a 5-Fr catheter, a 14-mm long stainless steel tube for generating the pulsed water jet, a nozzle and an optical quartz fiber. The pulsed water jet was generated at pulse rates of 3 Hz by irradiating the physiological saline (4°C) within the stainless steel tube with an holmium-doped yttrium-aluminum-garnet (Ho:YAG) laser at 1.1 J/pulse. Mechanical characteristics were evaluated using a force meter. The device was used only for the part of submucosal dissection in the swine ESD model. Tissues removed using the pulsed jet device and a conventional electrocautery device, and the esophagus, were histologically examined to assess thermal damage.

Results: The peak impact force was observed at a stand-off distance of 40 mm (1.1 J/pulse). ESD using the pulsed jet device was successful, as the tissue specimens showed precise dissection of the submucosal layer. The extent of thermal injury was significantly lower in the dissected bed using the pulsed jet device.

Conclusion: The results showed that the present endoscopic pulsed jet system is a useful alternative for a safe ESD with minimum tissue injury.

Download full-text PDF

Source
http://dx.doi.org/10.1111/j.1443-1661.2012.01375.xDOI Listing

Publication Analysis

Top Keywords

pulsed jet
24
jet device
16
water jet
12
submucosal dissection
12
pulsed
9
jet
9
endoscopic submucosal
8
stainless steel
8
steel tube
8
pulsed water
8

Similar Publications

Evaluation of the Digital Ventilated Cage® system for circadian phenotyping.

Sci Rep

January 2025

Sir Jules Thorn Sleep and Circadian Neuroscience Institute, Kavli Institute for Nanoscience Discovery, Nuffield Department of Clinical Neurosciences, University of Oxford, Dorothy Crowfoot Hodgkin Building, South Parks Road, Oxford, OX1 3QU, UK.

The study of circadian rhythms has been critically dependent upon analysing mouse home cage activity, typically employing wheel running activity under different lighting conditions. Here we assess a novel method, the Digital Ventilated Cage (DVC, Tecniplast SpA, Italy), for circadian phenotyping. Based upon capacitive sensors mounted under black individually ventilated cages with inbuilt LED lighting, each cage becomes an independent light-controlled chamber.

View Article and Find Full Text PDF

High resolution infrared spectra of water-CO dimers are further studied using tunable infrared sources to probe a pulsed slit jet supersonic expansion. The relatively weak transition of DO-CO in the DO ν fundamental region (≈2760 cm) is observed for the first time, as are various spectra of DO-CO. Combination bands involving the intermolecular in plane geared bend (disrotatory) mode are observed for HO-CO (≈1642, 2397 cm) in the HO ν and CO ν regions, for HDO-CO (≈2761 cm) in the HDO ν region, and for DO-CO (≈2386, 2705 and 2821 cm) in the CO ν, DO ν, and DO ν regions.

View Article and Find Full Text PDF

The role of the hierarchical organization of the suprachiasmatic nucleus (SCN) in its functioning, jet lag, and the light treatment of jet lag remains poorly understood. Using the core-shell model, we mimic collective behavior of the core and shell populations of the SCN oscillators in transient states after rapid traveling east and west. The existence of a special region of slow dynamical states of the SCN oscillators can explain phenomena such as the east-west asymmetry of jet lag, instances when entrainment to an advance is via delay shifts, and the dynamics of jet lag recovery time.

View Article and Find Full Text PDF

Outcomes of a transapical edge-to-edge repair system in secondary mitral regurgitation.

J Invasive Cardiol

January 2025

Department of Cardiology, Zhongshan Hospital, Fudan University, Shanghai Institute of Cardiovascular Diseases, China; State Key Laboratory of Cardiovascular Diseases, Zhongshan Hospital, Fudan University; NHC Key Laboratory of Ischemic Heart Diseases; Key Laboratory of Viral Heart Diseases, Chinese Academy of Medical Sciences; National Clinical Research Center for Interventional Medicine, Shanghai, China.

Objectives: The ValveClamp system (Hanyu Medical Technology) is a novel transcatheter edge-to-edge repair (TEER) system designed for ease of operation; however, there is a lack of data on its application in secondary mitral regurgitation (SMR). The authors report the mid-term outcomes of TEER using the ValveClamp system in SMR.

Methods: The study prospectively analyzed consecutive severe SMR patients who underwent transapical ValveClamp implantation at 10 Chinese centers.

View Article and Find Full Text PDF

This Letter reports what we believe to be a novel schlieren approach with adaptive temporal resolution. The fundamental concept of this approach is to fuse an event-based camera and a low-speed frame-based camera to generate high-frame-rate videos by leveraging the strengths of both. Using a novel experimental setup, events and frames are accurately aligned in both space and time.

View Article and Find Full Text PDF

Want AI Summaries of new PubMed Abstracts delivered to your In-box?

Enter search terms and have AI summaries delivered each week - change queries or unsubscribe any time!