Several methods exist for the measurement of cyanide levels in treated wastewater,typically requiring preservation of the sample with sodium hydroxide to minimize loss of hydrogen cyanide gas (HCN). Recent reports have shown that cyanide levels may increase with chlorination or preservation. In this study, three flow injection analysis methods involving colorimetric and amperometric detection were compared within one laboratory, as well as across separate laboratories and equipment. Split wastewater samples from eight facilities and three different sampling periods were tested. An interlaboratory confidence interval of 3.5 ppb was calculated compared with the intralaboratory reporting limit of 2 ppb. The results show that free cyanide measurements are not statistically different than total cyanide levels. An artificial increase in cyanide level is observed with all methods for preserved samples relative to nonpreserved samples, with an average increase of 2.3 ppb. The possible loss of cyanide without preservation is shown to be statistically insignificant if properly stored up to 48 hours. The cyanide increase with preservation is further substantiated with the method of standard additions and is not a matrix interference. The increase appears to be correlated with the amount of cyanide observed without preservation, which appears to be greater in those facilities that disinfect their wastewater with chlorine, followed by dechlorination with sodium bisulfite.