Purpose: Dosimetric accuracy in radiation treatment of breast cancer is critical for the evaluation of cosmetic outcomes and survival. It is often considered that treatment planning systems (TPS) may not be able to provide accurate dosimetry in the buildup region. This was investigated in various treatment techniques such as tangential wedges, field-in-field (FF), electronic compensator (eComp), and intensity-modulated radiotherapy (IMRT).

Methods: Under Institutional Review Board (IRB) exemption, radiotherapy treatment plans of 111 cases were retrospectively analyzed. The distance between skin surface and 95% isodose line was measured. For measurements, Gafchromic EBT2 films were used on a humanoid unsliced phantom. Multiple layers of variable thickness of superflab bolus were placed on the breast phantom and CT scanned for planning. Treatment plans were generated using four techniques with two different grid sizes (1 × 1 and 2.5 × 2.5 mm(2)) to provide optimum dose distribution. Films were placed at different depths and exposed with the selected techniques. A calibration curve for dose versus pixel values was also generated on the same day as the phantom measurement was conducted. The DICOM RT image, dose, and plan data were imported to the in-house software. On axial plane of CT slices, curves were drawn at the position where EBT2 films were placed, and the dose profiles on the lines were acquired. The calculated and measured dose profiles were separated by check points which were marked on the films before irradiation. The segments of calculated profiles were stretched to match their resolutions to that of film dosimetry.

Results: On review of treatment plans, the distance between skin and 95% prescribed dose was up to 8 mm for plans of 27 patients. The film measurement revealed that the medial region of phantom surface received a mere 45%-50% of prescribed dose. For wedges, FF, and eComp techniques, region around the nipple received approximately 80% of prescribed dose, although only IMRT showed inhomogeneous dose profile. At deeper depths mainly (6-11 mm depths), film dosimetry showed good agreement with the TPS calculation. In contrast, the measured dose at a 3-mm depth was higher than TPS calculation by 15%-30% for all techniques. For the tangential and IMRT techniques, 1 × 1 mm(2) grid size showed a smaller difference than that with a 2.5 × 2.5 mm(2) grid size compared to the measurements.

Conclusions: In general, TPS even with advanced algorithms do not provide accurate dosimetry in the buildup region, as verified by EBT2 film for all treatment techniques. For all cases, TPS and measured doses were in agreement from 6 mm in depth but differed at shallower depths. Grid size plays an important role in dose calculation. For accurate dosimetry small grid size should be used where differences are lower between TPS and measurements.

Download full-text PDF

Source
http://dx.doi.org/10.1118/1.4770285DOI Listing

Publication Analysis

Top Keywords

grid size
16
treatment techniques
12
accurate dosimetry
12
treatment plans
12
prescribed dose
12
dose
11
treatment
9
treatment planning
8
breast cancer
8
techniques
8

Similar Publications

Objective Applying carbon ion beams, which have high linear energy transfer and low scatter within the human body, to Spatially Fractionated Radiation Therapy (SFRT) could benefit the treatment of deep-seated or radioresistant tumors. This study aims to simulate the dose distributions of spatially fractionated beams (SFB) to accurately determine the delivered dose and model the cell survival rate following SFB irradiation. Approach Dose distributions of carbon ion beams are calculated using the Triple Gaussian Model.

View Article and Find Full Text PDF

The principle of independence is a fundamental yet often disregarded assumption in statistical inference. It is observed that the implications of correlations, if not considered, can lead to a conservative estimation of Type I error in the presence of positive linear correlations when utilizing the Kolmogorov-Smirnov (KS) test. Conversely, negative linear correlations may engender a liberal estimation of Type I error.

View Article and Find Full Text PDF

Objective Applying carbon ion beams, which have high linear energy transfer and low scatter within the human body, to Spatially Fractionated Radiation Therapy (SFRT) could benefit the treatment of deep-seated or radioresistant tumors. This study aims to simulate the dose distributions of spatially fractionated beams (SFB) to accurately determine the delivered dose and model the cell survival rate following SFB irradiation. Approach Dose distributions of carbon ion beams are calculated using the Triple Gaussian Model.

View Article and Find Full Text PDF

This study addresses the thermal management challenge in battery systems by enhancing phase change material composites with Ni-P and Ni-P-Cu coatings on phase change material/expanded graphite structures. Traditional phase change materials are limited by low thermal conductivity and mechanical stability, which restricts their effectiveness in high-demand applications. Unlike previous studies, this work integrates Ni-P and Ni-P-Cu coatings to significantly improve both the thermal conductivity and mechanical strength of phase change material/expanded graphite composites, filling a crucial gap in battery thermal management solutions.

View Article and Find Full Text PDF

Detecting bolt defects on transmission lines is crucial for ensuring the safe operation of the electrical power system. However, existing methods for detecting bolt defects on transmission lines require higher detection accuracy and smaller model sizes. To address these challenges, this paper proposes a real-time bolt defect detection model based on YOLOv7, named YOLOv7-CWFD.

View Article and Find Full Text PDF

Want AI Summaries of new PubMed Abstracts delivered to your In-box?

Enter search terms and have AI summaries delivered each week - change queries or unsubscribe any time!