Background: Stress hyperglycemia in the critically ill is associated with increased morbidity and mortality. Continuous glucose monitoring offers a solution to the difficulties of dosing intravenous insulin properly to maintain glycemic control. The purpose of this study was to evaluate an intravascular continuous glucose monitoring (IV-CGM) system with a sensing element based on the concept of quenched fluorescence.
Method: A second-generation intravascular continuous glucose sensor was evaluated in 13 volunteer subjects with type 1 diabetes mellitus. There were 21 study sessions of up to 24 h in duration. Sensors were inserted into peripheral veins of the upper extremity, up to two sensors per subject per study session. Sensor output was compared with temporally correlated reference measurements obtained from venous samples on a laboratory glucose analyzer.
Results: Data were obtained from 23 sensors in 13 study sessions with 942 paired reference values. Fourteen out of 23 sensors (60.9%) had a mean absolute relative difference ≤ 10%. Eighty-nine percent of paired points were in the clinically accurate A zone of the Clarke error grid and met ISO 15197 performance criteria. Adequate venous blood flow was identified as a necessary condition for accuracy when local sensor readings are compared with venous blood glucose.
Conclusions: The IV-CGM system was capable of achieving a high level of glucose measurement accuracy. However, superficial peripheral veins may not provide adequate blood flow for reliable indwelling blood glucose monitoring.
Download full-text PDF |
Source |
---|---|
http://www.ncbi.nlm.nih.gov/pmc/articles/PMC3570865 | PMC |
http://dx.doi.org/10.1177/193229681200600604 | DOI Listing |
Anal Chim Acta
May 2025
College of Textile Science and Engineering, Jiangnan University, 1800Lihu Road, Wuxi, 214122, China. Electronic address:
Flexible sweat sensors possess the special potential for continuous non-invasive monitoring of human blood glucose. We put forward a flexible microcolumn array sensor, which is designed for health monitoring by means of detecting glucose levels in sweat and capturing physiological signals related to human movement. With the combination of silk fibroin (SF), waterborne polyurethane (PU), and multi-walled carbon nanotubes (MWCNT), this microcolumn film electrode is able to effectively function as a strain sensor benefiting from the superior mechanical performance of PU.
View Article and Find Full Text PDFBiol Pharm Bull
March 2025
Department of Pharmacy, College of Pharmacy, Kinjo Gakuin University, 2-1723 Omori, Moriyama-ku, Nagoya 463-8521, Japan.
We examined whether the glucose levels and awareness of individuals without diabetes changed after using a sensor-based intermittently scanned continuous glucose monitoring (isCGM) system in their daily lives. Japanese individuals without a diabetes diagnosis wore the isCGM system while maintaining a normal lifestyle during the baseline period. A certified diabetes educator coached them on how to improve their lifestyle based on information from sensor data, food journals, and body composition.
View Article and Find Full Text PDFJ Nutr
March 2025
Frontier Science Research Center, University of Miyazaki, Miyazaki 889-1692, Japan. Electronic address:
Background: Type 2 diabetes (T2D) is characterized by insulin resistance and defective insulin secretion. Previously, we found that rats fed soft pellets (SPs) on a 3-hour restricted schedule over 14 weeks demonstrated glucose intolerance and insulin resistance with disruption of insulin signaling.
Objective: To determine (1) the time required for an SP diet to induce insulin resistance, and (2) whether the metabolic derangements in rats fed SPs can be reversed by changing to a standard control diet.
Diabetes Metab J
March 2025
Division of Endocrinology and Metabolism, Department of Internal Medicine, St. Vincent's Hospital, College of Medicine, The Catholic University of Korea, Seoul, Korea.
Backgruound: Diabetes in older adults is becoming a significant public burden to South Korea. However, a comprehensive understanding of epidemiologic trends and the detailed clinical characteristics of older adults with diabetes is lacking. Therefore, we evaluated epidemiologic trends and the metabolic and lifestyle characteristics of diabetes in Korean older adults.
View Article and Find Full Text PDFArch Gerontol Geriatr
March 2025
School of Public Health, Capital Medical University, NO. 10 Xitoutiao, Youanmenwai, Fengtai District, Beijing 100069, China. Electronic address:
Background: Insulin resistance (IR) is strongly related to non-alcoholic fatty liver disease (NAFLD). Triglyceride-glucose (TyG) index serves as a novel substitute indicator for IR. However, research on the effect of TyG index on NAFLD remains sparse.
View Article and Find Full Text PDFEnter search terms and have AI summaries delivered each week - change queries or unsubscribe any time!