Purpose: The aims of this study were to evaluate the incidence of anchor penetration of the far cortex of the glenoid neck after arthroscopic Bankart repair and to compare the biomechanical properties of anchors in the 4- and 5:30- to 6-o'clock positions on the glenoid.
Methods: Twelve (6 matched pairs) fresh-frozen human cadaveric shoulders were used to simulate arthroscopic Bankart repair in the lateral decubitus position. The most inferior anchor (5:30 to 6 o'clock) and that above it (4 o'clock) were inserted via the anteroinferior portal on the glenoid using the standard technique. After both anchor insertions, anchor perforation of the glenoid far cortex was identified. Biomechanical properties were measured to determine cyclic displacement of anchors at 100 and 500 cycles, stiffness, yield load, and ultimate failure strength.
Results: All 12 suture anchors (100%) at 5:30 to 6 o'clock penetrated throughout the far cortex, whereas only 4 anchors (33%) at 4 o'clock did so (P = .005). The mean distance the anchor tip traveled into far cortex was significantly longer at 5:30 to 6 o'clock than at 4 o'clock (6.8 ± 1.6 mm v 2.0 ± 1.6 mm, P = .001). In terms of mechanical strength, anchors at 5:30 to 6 o'clock had greater 100- and 500-cycle mean displacements than those at 4 o'clock (3.0 ± 0.5 mm v 2.5 ± 0.3 mm, P = .018 for 100 cycles; 3.5 ± 0.7 mm v 2.8 ± 0.3 mm, P = .018 for 500 cycles), although no differences in ultimate failure strength after cyclic loading were found between 2 positions (133.4 ± 40.3 and 133.7 ± 29.2 N, respectively; P = .985).
Conclusions: For arthroscopic Bankart repair, insertion of the most inferior anchor via the anteroinferior portal with standard technique, in the lateral decubitus position, carries a high risk of perforating the inferior far cortex of the glenoid (100% in our study). This may result in mechanical weakness of the most inferior repair specifically in the early postoperative period.
Clinical Relevance: Perforation of the glenoid far cortex by the most inferior anchor and its mechanical weakness should be taken into consideration. Further study is needed to improve surgical technique to place the most inferior anchor in an optimal position by arthroscopy.
Download full-text PDF |
Source |
---|---|
http://dx.doi.org/10.1016/j.arthro.2012.08.013 | DOI Listing |
Laryngoscope Investig Otolaryngol
February 2025
Department of Otolaryngology-Head and Neck Surgery, Dentistry and Pharmaceutical Sciences Okayama University Graduate School of Medicine Okayama Japan.
Objectives: Although various stapedotomy and stapedectomy techniques exist, anchoring the piston can be challenging. We present a novel surgical approach for treating congenital stapes malformations with an atypical facial nerve trajectory.
Methods: This is a case of a 7-year-old boy presenting with bilateral conductive hearing loss.
Background: Failure after rotator cuff repair is typically due to a loss of integrity of the bone-tendon interface. The BioWick anchor (Zimmer-Biomet) is an interpositional scaffold-anchor that was developed to improve tendon-bone healing. The purpose of this study was to determine the clinical efficacy of this novel anchor compared with a standard anchor with respect to retear rates and patient outcomes.
View Article and Find Full Text PDFEur Spine J
December 2024
Department of Orthopaedics, Jingjiang People's Hospital Affiliated to Yangzhou University, Jingjiang, Taizhou, Jiangsu Province, 214500, China.
Objective: For anterior cervical discectomy and fusion (ACDF), stand-alone anchored spacers (SAAS) and cage-plate system (CPS) are currently employed. However, controversy remains over the effectiveness and security of these two apparatuses in multilevel ACDF. The aim of this study was to demonstrate the global long-term effectiveness and safety of SAAS versus CPS with multilevel ACDF.
View Article and Find Full Text PDFOrthop J Sports Med
December 2024
Department of Trauma, Hand and Reconstructive Surgery, University Hospital Münster, Münster, Germany.
Background: Several types of suture anchors, which differ in their working principles, are available for fixation of ligamentous structures in knee surgery. How the choice of a suture anchor type influences the biomechanical stability of ligament fixation is largely unknown.
Purpose: To compare the biomechanical properties of different suture anchor designs regarding primary stability for tendon fixation and repair in medial collateral ligament (MCL) surgery.
J Affect Disord
December 2024
Dept. Psychology, New Zealand. Electronic address:
Purpose: Anxiety disorders are a major global issue. Diagnosis via symptoms, not biological causes, delivers poor treatment outcomes. Our frontal EEG biomarker, Goal Conflict Specific Rhythmicity (GCSR; 4-12 Hz), developed from our long-standing detailed neuropsychological theory of anxiety processes, is reduced by all chemical types of selective anxiolytic and is high in cases across a range of currently diagnosed anxiety disorders.
View Article and Find Full Text PDFEnter search terms and have AI summaries delivered each week - change queries or unsubscribe any time!