Aims: Adoption of fractional flow reserve (FFR) remains low (6-8%), partly because of the time, cost and potential inconvenience associated with vasodilator administration. The instantaneous wave-Free Ratio (iFR) is a pressure-only index of stenosis severity calculated without vasodilator drugs. Before outcome trials test iFR as a sole guide to revascularisation, we evaluate the merits of a hybrid iFR-FFR decision-making strategy for universal physiological assessment.
Methods And Results: Coronary pressure traces from 577 stenoses were analysed. iFR was calculated as the ratio between Pd and Pa in the resting diastolic wave-free window. A hybrid iFR-FFR strategy was evaluated, by allowing iFR to defer some stenoses (where negative predictive value is high) and treat others (where positive predictive value is high), with adenosine being given only to patients with iFR in between those values. For the most recent fixed FFR cut-off (0.8), an iFR of <0.86 could be used to confirm treatment (PPV of 92%), whilst an iFR value of >0.93 could be used to defer revascularisation (NPV of 91%). Limiting vasodilator drugs to cases with iFR values between 0.86 to 0.93 would obviate the need for vasodilator drugs in 57% of patients, whilst maintaining 95% agreement with an FFR-only strategy. If the 0.75-0.8 FFR grey zone is accounted for, vasodilator drug requirement would decrease by 76%.
Conclusion: A hybrid iFR-FFR decision-making strategy for revascularisation could increase adoption of physiology-guided PCI, by more than halving the need for vasodilator administration, whilst maintaining high classification agreement with an FFR-only strategy.
Download full-text PDF |
Source |
---|---|
http://dx.doi.org/10.4244/EIJV8I10A179 | DOI Listing |
Int J Numer Method Biomed Eng
October 2020
Center of Excellence in Energy Conversion, School of Mechanical Engineering, Sharif University of Technology, Tehran, Iran.
The objective of this paper is to apply computational fluid dynamic (CFD) as a complementary tool for clinical tests to not only predict the present and future status of left coronary artery stenosis but also to evaluate some clinical hypotheses. In order to assess the present status of the coronary artery stenosis severity, and thereby selecting the most appropriate type of treatment for each patient, fractional flow reserve (FFR), instantaneous wave free-ratio (iFR), and coronary flow reserve (CFR) are calculated. To examine FFR, iFR, and CFR results, the effect of geometric features of stenoses, including diameter reduction (%), lesion length (LL), and minimum lumen diameter (MLD), is studied on them.
View Article and Find Full Text PDFCardiovasc Revasc Med
July 2019
Division of Cardiology, Department of Medicine, University of Verona, Italy. Electronic address:
Background: The complexity of coronary physiology in presence of severe aortic stenosis (AS) raises concerns about the reliability of pressure-derived indexes in this clinical setting. Furthermore, neither fractional flow reserve (FFR) nor instantaneous wave-free period (iFR) has been validated in AS. Combining iFR and FFR in a tailored decision-making strategy may help to increase simplicity, accuracy and safety of physiology-guided revascularization in AS.
View Article and Find Full Text PDFJ Interv Cardiol
December 2017
Department of Cardiology, Liverpool Heart and Chest Hospital, Liverpool, UK.
Aims: To evaluate the correlation between iFR and FFR in real-world clinical practice.
Methods And Results: Retrospective, single-centre study of 229 consecutive pressure-wire studies (n = 158). Real-time iFR and FFR measurements were performed for angiographically borderline stenoses.
Front Cardiovasc Med
May 2017
Cardiac Unit, Raigmore Hospital, Inverness, UK.
Background: The instantaneous wave-free ratio (iFR) is a novel method to assess the ischemic potential of coronary artery stenoses. Clinical trial data have shown that iFR has acceptable diagnostic agreement with fractional flow reserve (FFR), the reference standard for the functional assessment of coronary stenoses. This study compares iFR measurements with FFR measurements in a real world, single-center setting.
View Article and Find Full Text PDFCirc Cardiovasc Interv
November 2016
From the University of Glasgow, United Kingdom (B.H., K.G.O., C.B., P.M., M.B.M., H.E., M.C.P., P.R., R.G., M.M.L., S.H., S.W.); Cardiology Department, Golden Jubilee National Hospital, Clydebank, Glasgow, United Kingdom (B.H., K.G.O., C.B., J.M.); and The Weatherhead PET Imaging Center, Houston, TX (N.J.).
Background: Distal coronary to aortic pressure ratio (Pd/Pa) and instantaneous wave-free ratio (iFR) are indices of functional significance of a coronary stenosis measured without hyperemia. It has been suggested that iFR has superior diagnostic accuracy to Pd/Pa when compared with fractional flow reserve (FFR).We hypothesized that in comparison with FFR, revascularization decisions based on either binary cutoff values for iFR and Pd/Pa or hybrid strategies incorporating iFR or Pd/Pa will result in similar levels of disagreement.