Background: Inhalation of low-density helium/oxygen mixtures has been used both to lower the airway resistance and work of breathing of patients with obstructive lung disease and to transport pharmaceutical aerosols to obstructed lung regions. However, recent clinical investigations have highlighted the potential for entrainment of room air to dilute helium/oxygen mixtures delivered through non-rebreather facemasks, thereby increasing the density of the inhaled gas mixture and limiting intended therapeutic effects. This article describes the development of benchtop methods using face models for evaluating delivery of helium/oxygen mixtures through facemasks.

Methods: Four face models were used: a flat plate, a glass head manikin, and two face manikins normally used in life support training. A mechanical test lung and ventilator were employed to simulate spontaneous breathing during delivery of 78/22 %vol helium/oxygen through non-rebreather facemasks. Based on comparison of inhaled helium concentrations with available clinical data, one face model was selected for measurements made during delivery of 78/22 or 65/35 %vol helium/oxygen through three different masks as tidal volume varied between 500 and 750 ml, respiratory rate between 14 and 30 breaths/min, the inspiratory/expiratory ratio between 1/2 and 1/1, and the supply gas flow rate between 4 and 15 l/min. Inhaled helium concentrations were measured both with a thermal conductivity analyzer and using a novel flow resistance method.

Results: Face models borrowed from life support training provided reasonably good agreement with available clinical data. After normalizing for the concentration of helium in the supply gas, no difference was noted in the extent of room air entrainment when delivering 78/22 versus 65/35 %vol helium/oxygen. For a given mask fitted to the face in a reproducible manner, delivered helium concentrations were primarily determined by the ratio of supply gas flow rate to simulated patient minute ventilation, with the inspiratory/expiratory ratio playing a secondary role. However, the functional dependence of helium concentration on these two ratios depended on the mask design.

Conclusions: Large differences in mask performance were identified. With continued refinement, the availability of reliable benchtop methods is expected to assist in the development and selection of patient interfaces for delivery of helium/oxygen and other medical gases.

Download full-text PDF

Source
http://www.ncbi.nlm.nih.gov/pmc/articles/PMC3557179PMC
http://dx.doi.org/10.1186/2045-9912-2-31DOI Listing

Publication Analysis

Top Keywords

non-rebreather facemasks
12
helium/oxygen mixtures
12
face models
12
%vol helium/oxygen
12
helium concentrations
12
supply gas
12
helium/oxygen
8
room air
8
benchtop methods
8
delivery helium/oxygen
8

Similar Publications

To evaluate aerosol exposure risk and prevention strategies during bystander, pre-hospital, and inpatient cardiopulmonary resuscitation (CPR). This study compared hands-only CPR, CPR with a surgical or N95 mask, and CPR with a non-rebreather mask at 15 L/min. 30:2 compression-ventilation ratio CPR was tested with face-mask ventilation (FMV), FMV with a high efficiency particulate air (HEPA) filter; supraglottic airway (SGA), SGA with a surgical mask, SGA with a HEPA filter, or SGA with both.

View Article and Find Full Text PDF

Introduction: Because of the COVID-19 pandemic, intensive care units (ICU) can be overwhelmed by the number of hypoxemic patients.

Material And Methods: This single centre retrospective observational cohort study took place in a French hospital where the number of patients exceeded the ICU capacity despite an increase from 18 to 32 beds. Because of this, 59 (37%) of the 159 patients requiring ICU care were referred to other hospitals.

View Article and Find Full Text PDF

Analysis of the airway registry from an academic emergency department in South Africa.

S Afr Med J

May 2020

Division of Emergency Medicine, Faculty of Health Sciences, University of the Witwatersrand, Johannesburg, South Africa.

Background: Airway management is an essential skill for doctors working in the emergency department (ED). Safety and efficacy are crucial to its success. Analysis of an airway registry can provide feedback that can be used for quality improvement purposes.

View Article and Find Full Text PDF

The aim of this study was to test the effects of CPAP on moderately sedated patients undergoing colonoscopy. Our hypothesis was that CPAP can reduce the incidence and duration of obstructive apnea and hemoglobin oxygen desaturation in patients undergoing procedural sedation for colonoscopy. Two groups of consenting adult patients scheduled to undergo routine colonoscopy procedures and sedated with propofol and fentanyl were monitored in this study: control and intervention.

View Article and Find Full Text PDF

Background: Inhalation of low-density helium/oxygen mixtures has been used both to lower the airway resistance and work of breathing of patients with obstructive lung disease and to transport pharmaceutical aerosols to obstructed lung regions. However, recent clinical investigations have highlighted the potential for entrainment of room air to dilute helium/oxygen mixtures delivered through non-rebreather facemasks, thereby increasing the density of the inhaled gas mixture and limiting intended therapeutic effects. This article describes the development of benchtop methods using face models for evaluating delivery of helium/oxygen mixtures through facemasks.

View Article and Find Full Text PDF

Want AI Summaries of new PubMed Abstracts delivered to your In-box?

Enter search terms and have AI summaries delivered each week - change queries or unsubscribe any time!