Study Design:  Systematic review.Clinical questions: (1) What evidence is available from studies of cervical total disc arthroplasty (C-ADR) failure and retrieval regarding durability, wear, and reasons for failure of C-ADR? (2) What evidence is available from experimental models regarding the durability of C-ADR beyond 5 years?

Methods:  We searched electronic databases to identify published reports of explanted cervical artificial discs and biomechanical simulations of disc wear.

Results:  Nine articles were identified describing 17 devices explanted from human patients and four articles describing 23 devices explanted from non-human subjects. Wear properties were not consistently reported across studies, so summaries for specific variables are based on few cases. No device had been implanted longer than 4 years. In both human and non-human subjects, devices showed evidence of metallic and polymeric (for discs with polymer components) debris. Inflammatory cells were frequently present in surrounding soft tissues. Signs of infection were uncommon.Four patients had reactions interpreted as hypersensitivity to metal. We identified three articles on biomechanical wear simulations. Devices were tested between 10 and 20 million cycles in axial loading, flexion/extension, and lateral bending. No device failures were reported. One study suggests such simulations may represent 50 or more years of wear in actual patients.

Conclusion:  Cervical disc implants consistently produced polymeric and metallic debris, which was typically accompanied by inflammation. Hypersensitivity to metal may increase risk for device failure. Biomechanical simulations indicate that cervical disc implants may be durable beyond the currently reported length of clinical follow-up.

Download full-text PDF

Source
http://www.ncbi.nlm.nih.gov/pmc/articles/PMC3519402PMC
http://dx.doi.org/10.1055/s-0031-1298606DOI Listing

Publication Analysis

Top Keywords

cervical artificial
8
biomechanical simulations
8
describing devices
8
devices explanted
8
non-human subjects
8
hypersensitivity metal
8
disc implants
8
disc
5
wear
5
systematic review
4

Similar Publications

Purpose: Human epidermal growth factor receptor 2 (HER2)-targeted therapies have shown promise in treating -amplified metastatic colorectal cancer (mCRC). Identifying optimal biomarkers for treatment decisions remains challenging. This study explores the potential of artificial intelligence (AI) in predicting treatment responses to trastuzumab plus pertuzumab (TP) in patients with -amplified mCRC from the phase II TRIUMPH trial.

View Article and Find Full Text PDF

Objective: This study aims to evaluate the impact of prompt construction on the quality of artificial intelligence (AI) chatbot responses in the context of head and neck surgery.

Study Design: Observational and evaluative study.

Setting: An international collaboration involving 16 researchers from 11 European centers specializing in head and neck surgery.

View Article and Find Full Text PDF

Nerve conduction F-wave studies contain critical information about subclinical motor dysfunction which may be used to diagnose patients with amyotrophic lateral sclerosis (ALS). However, F-wave responses are highly variable in morphology, making waveform interpretation challenging. Artificial Intelligence techniques can extract time-frequency features to provide new insights into ALS diagnosis and prognosis.

View Article and Find Full Text PDF

Background: The objective of this report is to present a case of two cervical spine artificial discs (Bryan Cervical Disk) that completely disappeared within 6 months as a result of a high-energy trauma more than 10 years after the initial surgery. Implant dislocation is a known complication in artificial cervical disc replacement. However, this report presents the case of an exceptional migration path with esophageal ingrowth and rectal excretion, not only for one artificial disc but for two at different times It highlights the need for long-term follow-up examinations after artificial cervical disc arthroplasty (ACDA).

View Article and Find Full Text PDF

Want AI Summaries of new PubMed Abstracts delivered to your In-box?

Enter search terms and have AI summaries delivered each week - change queries or unsubscribe any time!