Tidal volume delivery from ICU ventilators at BTPS conditions: a bench study.

Respir Care

Haute Etude d'Ingénieurs, Lille, France.

Published: April 2013

Background: Even though it is not a common practice, an external filter to the expiratory limb of the breathing circuit may protect the expiratory valve from water saturation in case of nebulization, or from the environment in case of lung infection with multi-drug-resistant micro-organisms or H1N1 influenza.

Methods: We added an external filter to the expiratory limb and measured tidal volume (VT) from 6 ICU ventilators: 2 with built-in expiratory filter (Avea, Puritan Bennett 840), and 4 without (Engström Carestation, Evita XL, Evita V500, and Servo-i), set in volume controlled mode, at BTPS (body temperature and pressure saturated) condition, with a heated humidifier and a lung model (compliance 16 mL/cm H2O, resistance 20 cm H2O/L/s) placed inside a neonatal incubator. The temperature was targeted at 37°C for both the heated humidifier and the incubator. The setup was run continuously for 24 hours. In the latter 4 ICU ventilators, a Hygrobac or Sterivent S external filter was placed upstream from the expiratory valve for an additional 24-hour period for each. At the end of this period, VT was measured at 4 nominal VT values (300, 400, 500, and 800 mL) with a pneumotachograph. The volume error computed from the ratio of set to measured VT (% set VT) was the primary end point.

Results: In these warm and wet conditions, volume error averaged 96 ± 3% for Avea, 100 ± 7% for Puritan Bennett 840, 90 ± 2% for Evita XL, 100 ± 7% for Evita V500, 105 ± 2% for Servo-i, and 108 ± 4% for Engström Carestation (P < .001). With the Hygrobac the values were 93 ± 1% for Evita XL, 94 ± 4% for Evita V500, 110 ± 4% for Servo-i, and 99 ± 2% for Engström Carestation (P < .001). With the Sterivent the corresponding values were 95 ± 2%, 105 ± 2%, 112 ± 5%, and 98 ± 2%, respectively (P < .001).

Conclusions: In BTPS condition, volume error differed substantially across ICU ventilators for VT delivery, with further significant changes occurring after addition of a filter at the distal expiratory limb.

Download full-text PDF

Source
http://dx.doi.org/10.4187/respcare.01915DOI Listing

Publication Analysis

Top Keywords

icu ventilators
16
external filter
12
expiratory limb
12
engström carestation
12
evita v500
12
volume error
12
tidal volume
8
filter expiratory
8
expiratory valve
8
puritan bennett
8

Similar Publications

Sex Differences in Neurological Outcome at 6 and 12 Months Following Severe Traumatic Brain Injury. An Observational Analysis of the OXY-TC Trial.

J Neurotrauma

January 2025

Department of Anaesthesia and Intensive Care, Centre Hospitalier Universitaire Grenoble, and Inserm, U1216, Grenoble Institut Neurosciences, University Grenoble Alpes, Grenoble, France.

The effect of sex in outcomes after severe traumatic brain injury (TBI) remains uncertain. We explored whether outcomes differed between women and men after standardized care management during the first 5 days in the intensive care unit (ICU). This study was an observational analysis of the OXY-TC multicenter randomized clinical trial between June 15, 2016 and April 17, 2021.

View Article and Find Full Text PDF

The pulmonary contusion score: Development of a simple scoring system for blunt lung injury.

Surg Pract Sci

June 2024

Division of Acute Care Surgery, Department of Surgery, Kirk Kerkorian School of Medicine at UNLV, 1701W. Charleston Blvd, Suite 490, Las Vegas, NV 89102, United States.

Background: Pulmonary contusions (PC) are common after blunt chest trauma and can be identified with computed tomography (CT). Complex scoring systems for grading PC exist, however recent scoring systems rely on computer-generated algorithms that are not readily available at all hospitals. We developed a scoring system for grading PC to predict the need for prolonged mechanical ventilation and initial hospital admission location.

View Article and Find Full Text PDF

Objectives: This study aimed to investigate the association of baseline coagulation proteins with hospitalization variables in COVID-19 patients admitted to ICU, as well as coagulation system changes after one-year post-discharge, taking into account gender-specific bias in the coagulation profile.

Methods: We conducted a prospective longitudinal study on 49 ICU-admitted COVID-19 patients. Proteins were measured using a Luminex 200™.

View Article and Find Full Text PDF

Objective: Severe community-acquired pneumonia (sCAP) is one of the major diseases within the ICU. We hypothesize that subtyping sCAP based on simple inflammatory markers, organ dysfunction, and clinical metagenomics results is feasible.

Method: In this study, we retrospectively enrolled immunocompetent sCAP patients requiring invasive mechanical ventilation, who underwent clinical metagenomics from 17 medical centers.

View Article and Find Full Text PDF

Background: Subglottic stenosis is a significant clinical challenge in pediatric anesthesia, often necessitating interventions that can lead to various postoperative complications. The aim of this study was to determine the effect of prophylactic continuous positive airway pressure (CPAP) application on recovery time and airway complications in pediatric patients with subglottic stenosis undergoing balloon dilatation.

Methods: A prospective, double-blinded, parallel-group, randomized controlled study was conducted at Health Sciences University Ümraniye Training and Research Hospital on pediatric patients with subglottic stenosis, aged from 0 to 12 years and who underwent elective balloon dilatation under general anesthesia.

View Article and Find Full Text PDF

Want AI Summaries of new PubMed Abstracts delivered to your In-box?

Enter search terms and have AI summaries delivered each week - change queries or unsubscribe any time!