Sensing of left ventricular (LV) activity in some devices used for cardiac resynchronization therapy (CRT) was designed primarily to prevent the delivery of an LV stimulus into the LV vulnerable period. Such a sensing function of the LV channel is not universally available in contemporary CRT devices. Recordings of LV electrograms may provide special diagnostic data unavailable solely from the standard right ventricular electrogram and corresponding marker channel. We used the LV sensing function of Biotronik CRT defibrillators to find 3 cases of dissimilar ventricular rhythms or tachyarrhythmias. Such arrhythmias are potentially important because concomitant slower right ventricular activity may prevent or delay implantable cardioverter-defibrillator therapy for a life-threatening situation involving a faster and more serious LV tachyarrhythmia. Dissimilar ventricular rhythms may not be rare and may account for cases of unexplained sudden death with a normally functioning implantable cardioverter-defibrillator and no recorded terminal arrhythmia.
Download full-text PDF |
Source |
---|---|
http://dx.doi.org/10.1016/j.hrthm.2012.12.004 | DOI Listing |
Enter search terms and have AI summaries delivered each week - change queries or unsubscribe any time!